Interpretation:The type of noises which are frequency dependent or frequency independent needs to be determined.
Concept introduction:Chemical analyses are related with two types of noise which are chemical noise and instrumental noise. Chemical noise is due to reasons that affect the chemistry of the system being analyzed. Instrumental noise is components of the instrument. There are four types of instrumental noises, they are thermal noise, shot noise, flicker noise and environmental noise.
Answer to Problem 5.1QAP
Flicker noise and environmental noise are frequency dependent.
Thermal noise and shot noise are frequency independent.
Explanation of Solution
Thermal noise can be calculated according to below equation:
Here,
K = Boltzmann’s constant
T = Temperature
B = Band width
R= Resistance
According to above equation, thermal noise is dependent on band width but independent of frequency itself. Like thermal noise, shot noise is thus the same at any frequency.
Flicker noise is inversely proportional to the frequency being observed. It is frequency dependent.Environmental noise is a combination of different forms of noises that arise from the surroundings. Therefore, environmental noise is frequency dependent.
Here, it can be concluded that flicker noise and environmental noise are frequency dependent, while, thermal noise and shot noise are frequency independent according to above explanation.
Want to see more full solutions like this?
Chapter 5 Solutions
Principles of Instrumental Analysis, 6th Edition
- Please provide with answer, steps and explanation of ideas to solve.arrow_forwardIndicate whether the copper(II) acetate dimer, in its dihydrated form with the formula [(CH3COO)2Cu]2·2H2O, is a metal cluster, a cage compound, or neither.arrow_forwardPlease correct answer and don't use hand ratingarrow_forward
- Don't used hand raitingarrow_forwardReagan is doing an atomic absorption experiment that requires a set of zinc standards in the 0.4- 1.6 ppm range. A 1000 ppm Zn solution was prepared by dissolving the necessary amount of solid Zn(NO3)2 in water. The standards can be prepared by diluting the 1000 ppm Zn solution. Table 1 shows one possible set of serial dilutions (stepwise dilution of a solution) that Reagan could perform to make the necessary standards. Solution A was prepared by diluting 5.00 ml of the 1000 ppm Zn standard to 50.00 ml. Solutions C-E are called "calibration standards" because they will be used to calibrate the atomic absorption spectrometer. Table 1: Dilutions of Zinc Solutions Solution Zinc Solution Volume Diluted Solution Concentration used volume (ppm Zn) (mL) (mL) concentration (ppm Zn) Solution concentration A 1000 5.00 50.00 1.00×10² (ppm Zn(NO3)2) 2.90×10² Solution concentration (M Zn(NO3)2 1.53×10-3 B Solution A 5.00 100.00 5.00 C Solution B 5.00 50.00 0.50 7.65×10-6 D Solution B 10.00 50.00…arrow_forwardNonearrow_forward
- Principles of Instrumental AnalysisChemistryISBN:9781305577213Author:Douglas A. Skoog, F. James Holler, Stanley R. CrouchPublisher:Cengage LearningPhysical ChemistryChemistryISBN:9781133958437Author:Ball, David W. (david Warren), BAER, TomasPublisher:Wadsworth Cengage Learning,