(a)
Interpretation:
Length of a mirror drive is to be determined in Michelson interferometer in such way that the produced resolution will be enough for the separation of IR peaks at 600.7 cm-1 and 600.4 cm-1.
Concept introduction:
Resolution is calculated by the difference of the wave number of both IR peaks.
For the calculation of the length of the mirror, the following formula will be used:
Here,
(b)
Interpretation:
Length of a mirror drive is to be determined in Michelson interferometer in such a way that the produced resolution will be enough for the separation of IR peaks at 3002.5 cm-1 and 3008.6 cm-1.
Concept introduction:
Resolution is calculated by the difference in wave number of both IR peaks.
For the calculation of the length of the mirror, following formula will be used:
Here,
Trending nowThis is a popular solution!
Chapter 7 Solutions
Principles of Instrumental Analysis
- A student calibrated the mini-spectrophotometer with a blank solution, and then went to check their solution. They got a fingerprint on the cuvette while filling it. Because of this, the expected reading for the absorbance would be too high which would cause the calculated value of K, to be too higharrow_forwardA spectrophotometer has basic several components. Describe the role of each component and the justification for the setup of an instrument used for excitation spectroscopy and an instrument for emission spectroscopy.arrow_forwardI want clear handwritten solution only....i will up votearrow_forward
- P3arrow_forward3. In molecular absorption spectrophotometry, the relationship between transmittance and concentration is not linear while the absorbance vs concentration is linear. True Falsearrow_forwardAnswer the questions regarding a single Fourier transform infrared spectroscopy (FTIR) experiment. The interferogram was sampled at a retardation interval (8) of 0.0001297 cm. What wavenumber range is possible with a sample retardation interval (8) of 0.0001297 cm? Aỹ = cm- For a spectrum acquired in 8-bit space (256 data points sampled) for ô = -A to d = +A , determine the value of maximum retardation (A) and the resolution (R). A = cm R = cm-1 What is the interferometer mirror velocity if each data point required 267 us (tintt) to obtain? Vint = cm/sarrow_forward
- The percentage disodium fumarate at 250 nm and 25°C is 19.2% for a 5 x 10-4 M solution in a 1-cm cell. (a) Calculate the absorbance and transmittance of an aqueous solution of the molar absorption coefficient percentage transmittance of a 1.75 x 10-5 M solution in (b) What will be the a 10-cm cell? CS Scanned with CamScannerarrow_forwardThe points on the graph are used to depict the intensity of the ray detected. FT-IR spectrophotometers collect the interferograms of the sample and Fourier-transform them to an absorption spectrum and identify its band spectra analysis. Observe what you see in the graph.arrow_forwardAlexa Fluor 568 is a fluorescent dye that in aqueous solution displays a maximum absorbance at 576 nm, a quantum yield of 0.69, a lifetime of 3.60 ns and a Stokes shift of 24 nm. (i) Show a schematic of a typical fluorimeter and its main components. Briefly explain the role of each component. (ii) Explain what is the "Stokes shift" and what physical processes underpin the observation of such shifts; use diagrams and/or equations as appropriate in your explanation. (iii) Calculate the radiative decay constant of Alexa Fluor 568.arrow_forward
- Principles of Instrumental AnalysisChemistryISBN:9781305577213Author:Douglas A. Skoog, F. James Holler, Stanley R. CrouchPublisher:Cengage LearningPhysical ChemistryChemistryISBN:9781133958437Author:Ball, David W. (david Warren), BAER, TomasPublisher:Wadsworth Cengage Learning,
- Chemistry for Engineering StudentsChemistryISBN:9781285199023Author:Lawrence S. Brown, Tom HolmePublisher:Cengage LearningPrinciples of Modern ChemistryChemistryISBN:9781305079113Author:David W. Oxtoby, H. Pat Gillis, Laurie J. ButlerPublisher:Cengage Learning