Solution 1 0.8- Solution 2 0.6- 0.4- 0.2- 0.0 100 200 300 400 Wavelength (nm) UV-Visible spectroscopy was used to compare the two solutions (1 and 2). The first solution was green, whereas the second was yellow. The same axes were used to capture each solution throughout a variety of wavelengths. Above is an example of the resulting absorbance spectrum. (i). Ten (10) mL of solution 1 were mixed with 10 mL of solution 2 based on the UV-VIS spectrum. Describe the wavelength that should be used to measure the absorbance of solution 1 in this mixture. 1.0- 0.8- 0.6- 0.4- 2 0.2 - 0.0 4 Concentration (mM) Using two sets of standard solutions and blanks, the calibration curves for the two solutions were later obtained. As indicated in the graph above, the absorbances of both solutions were plotted on the same axis. When tested at the suitable wavelength, solution 1 showed a 0.4 absorbance. If solution 1 was chromium (III) sulfate, Cr2(SO4)3 explain the method of getting the concentration of the solution in mg L-1. [Mw Cr2(SO4)3 = 392.2 g mol-1] (iii). Both solutions have a concentration of around 3.5 x 10-3 M in a separate combination. Explain how UV-VIS spectroscopy may be used to reliably detect the concentrations of solution 1 and solution 2. Absorbance Absorbance

Chemistry
10th Edition
ISBN:9781305957404
Author:Steven S. Zumdahl, Susan A. Zumdahl, Donald J. DeCoste
Publisher:Steven S. Zumdahl, Susan A. Zumdahl, Donald J. DeCoste
Chapter1: Chemical Foundations
Section: Chapter Questions
Problem 1RQ: Define and explain the differences between the following terms. a. law and theory b. theory and...
icon
Related questions
Question

Kindly answer this question i, ii & iii

Solution 1
0.8-
Solution 2
0.6-
0.4-
0.2-
0.0 +
100
200
300
400
Wavelength (nm)
UV-Visible spectroscopy was used to compare the two solutions (1 and 2).
The first solution was green, whereas the second was yellow. The same axes
were used to capture each solution throughout a variety of wavelengths.
Above is an example of the resulting absorbance spectrum.
(i). Ten (10) mL of solution 1 were mixed with 10 mL of solution 2 based on
the UV-VIS spectrum. Describe the wavelength that should be used to
measure the absorbance of solution 1 in this mixture.
1.0 -
0.8-
0.6-
0.4-
0.2 -
0.0
Concentration (mM)
Using two sets of standard solutions and blanks, the calibration curves
for the two solutions were later obtained. As indicated in the graph
above, the absorbances of both solutions were plotted on the same
axis. When tested at the suitable wavelength, solution 1 showed a 0.4
absorbance. If solution 1 was chromium (III) sulfate, Cr2(SO4)3 explain
the method of getting the concentration of the solution in mg L-1. [Mw
Cr2(SO4)3 = 392.2 g mol-1]
(iii). Both solutions have a concentration of around 3.5 x 10-3 M in a separate
combination. Explain how UV-VIS spectroscopy may be used to reliably
detect the concentrations of solution 1 and solution 2.
Absorbance
Absorbance
Transcribed Image Text:Solution 1 0.8- Solution 2 0.6- 0.4- 0.2- 0.0 + 100 200 300 400 Wavelength (nm) UV-Visible spectroscopy was used to compare the two solutions (1 and 2). The first solution was green, whereas the second was yellow. The same axes were used to capture each solution throughout a variety of wavelengths. Above is an example of the resulting absorbance spectrum. (i). Ten (10) mL of solution 1 were mixed with 10 mL of solution 2 based on the UV-VIS spectrum. Describe the wavelength that should be used to measure the absorbance of solution 1 in this mixture. 1.0 - 0.8- 0.6- 0.4- 0.2 - 0.0 Concentration (mM) Using two sets of standard solutions and blanks, the calibration curves for the two solutions were later obtained. As indicated in the graph above, the absorbances of both solutions were plotted on the same axis. When tested at the suitable wavelength, solution 1 showed a 0.4 absorbance. If solution 1 was chromium (III) sulfate, Cr2(SO4)3 explain the method of getting the concentration of the solution in mg L-1. [Mw Cr2(SO4)3 = 392.2 g mol-1] (iii). Both solutions have a concentration of around 3.5 x 10-3 M in a separate combination. Explain how UV-VIS spectroscopy may be used to reliably detect the concentrations of solution 1 and solution 2. Absorbance Absorbance
Expert Solution
trending now

Trending now

This is a popular solution!

steps

Step by step

Solved in 4 steps with 3 images

Blurred answer
Knowledge Booster
Potentiometric Titrations
Learn more about
Need a deep-dive on the concept behind this application? Look no further. Learn more about this topic, chemistry and related others by exploring similar questions and additional content below.
Recommended textbooks for you
Chemistry
Chemistry
Chemistry
ISBN:
9781305957404
Author:
Steven S. Zumdahl, Susan A. Zumdahl, Donald J. DeCoste
Publisher:
Cengage Learning
Chemistry
Chemistry
Chemistry
ISBN:
9781259911156
Author:
Raymond Chang Dr., Jason Overby Professor
Publisher:
McGraw-Hill Education
Principles of Instrumental Analysis
Principles of Instrumental Analysis
Chemistry
ISBN:
9781305577213
Author:
Douglas A. Skoog, F. James Holler, Stanley R. Crouch
Publisher:
Cengage Learning
Organic Chemistry
Organic Chemistry
Chemistry
ISBN:
9780078021558
Author:
Janice Gorzynski Smith Dr.
Publisher:
McGraw-Hill Education
Chemistry: Principles and Reactions
Chemistry: Principles and Reactions
Chemistry
ISBN:
9781305079373
Author:
William L. Masterton, Cecile N. Hurley
Publisher:
Cengage Learning
Elementary Principles of Chemical Processes, Bind…
Elementary Principles of Chemical Processes, Bind…
Chemistry
ISBN:
9781118431221
Author:
Richard M. Felder, Ronald W. Rousseau, Lisa G. Bullard
Publisher:
WILEY