Lab2_Quantifying_Electrical_Energy_Handout
pdf
keyboard_arrow_up
School
University of Massachusetts, Lowell *
*We aren’t endorsed by this school
Course
3020
Subject
Mechanical Engineering
Date
Dec 6, 2023
Type
Pages
6
Uploaded by PrivateWater4873
Department of Mechanical Engineering
MECH.3020
University of Massachusetts Lowell
Laboratory 2
MECH. 3020
P a g e
1 | 6
MECH.3020
Laboratory 2
Quantifying Electrical Energy
Introduction
Measurement engineering is defined as the proper selection and use of instrumentation systems
for the purpose of experimentation, condition monitoring, and process control. One of the main
components of an instrumentation system is the transducer. The transducer is an analog device,
which is used to convert a change in a mechanical or thermal quantity to an electric signal that
can be quantified.
The purpose of this experiment is to explore methods used to quantify electrical energy.
Alternating Current (AC) and Direct Current (DC) signals will be generated with a function
generator and a ±20 volt DC power supply, respectively. A digital multimeter and digital
oscilloscope will be used to quantify various signals. The 3dB down point of the multimeter will
be determined, indicating that measurement systems have a limited accuracy range depending on
the frequency of the signal.
The use of analog filters will also be explored as a tool for limiting
the effects of signals outside the range of interest.
Pre-Lab Assignment
Pre-Lab assignments are to be done in your lab notebook following normal lab notebook
documentation procedures.
The Pre-Labs are not to be handed in but will be checked in the lab
to assure that they have been completed.
1.
Read chapter 1 and 3 - Wheeler and Ganji textbook 2
nd
edition
2.
From textbook -
Problem 3.1
- An amplifier produces an output of 5 V when the input is
5 μV.
What are the gain G
and the decibel gain G
dB
?
3.
From textbook -
Problem 3.2
- An amplifier has a gain of 60 dB.
If the input voltage is
3mV, what is the output voltage?
4.
OPTIONAL
–
From textbook -
Problem 3.13
- The output of a dc generator that produces
a maximum voltage of 90 V is to be attenuated to 10 V for input to a filter.
Specify values
of the resistors in an attenuation network such that the loading error of the voltage at the
output terminals of the generator is 0.1%.
The output impedance of the generator is 10 Ω
and the filter has an input impedance of 100 Ω.
5.
A multimeter measures an 8.0 volt 100 Hz sine wave. Determine the peak-peak voltage,
peak voltage, and period.
Department of Mechanical Engineering
MECH.3020
University of Massachusetts Lowell
Laboratory 2
MECH. 3020
P a g e
2 | 6
Assignment 1:
Initialize the oscilloscope Tektronix TBS1032B
The initialization of the oscilloscope is required to set the instrument to a point from which all
measurements can be referenced. The following procedure should be used when the oscilloscope
is used to quantify signals generated from transducers.
Procedure
1.
Turn the oscilloscope on. The power button is located on the top left side of the
oscilloscope.
2.
A previous configuration may be loaded. Load its default setup by pressing the
“Default
Setup” button. The message “Default setup recalled” will appear in the message bar. This
bar is located at the bottom of the display.
3.
The oscilloscope automatically applies to all signals an amplification factor of 10. This
factor has to be changed to 1. Within the Vertical System Control area, push the yellow
button “1” to bring the CH 1 setup menu up. Press the “Probe 10X Voltage” button. Then,
press the “Attenuation 10X” button and by using the “Multipurpose” knob, change the
attenuation fact
or to “1X”.
4.
Set the Vertical Scale to 5Volts/DIV
5.
Set the Horizontal Scale to 5ms.
The oscilloscope is now set to quantify electric signals. This initialization process should be
employed whenever the oscilloscope is used to quantify the output from a transducer.
Assignment 2: Quantifying Sinusoidal Waveforms
A digital multimeter can be used to measure current, resistance, and the voltage of a signal.
When an AC signal is recorded, the multimeter displays the root mean square (V
rms
) value which
can be used to obtain the peak (V
p
), and peak-peak (V
pp
) values of the wave. The oscilloscope is
capable of measuring voltage only. However, unlike the multimeter, the oscilloscope displays the
signal on a TFT-display allowing the period (T) and frequency (f) of the signal to be calculated.
Procedure
1.
Figure 1 shows the proper equipment configuration. The output of the function generator
will be the input to the oscilloscope and multimeter. Connect the function generator to the
oscilloscope’s
CH. 1
.
Function
Generator
Multimeter
Oscilloscope
CH. 1
Figure 1. Block diagram for Assignment 2
Department of Mechanical Engineering
MECH.3020
University of Massachusetts Lowell
Laboratory 2
MECH. 3020
P a g e
3 | 6
2.
Turn the function generator on. Set the function generator to output a 100 Hz sinewave at
a maximum amplitude.
IMPORTANT:
Whenever you generate a new signal, use the multimeter and
oscilloscope to check the magnitude and frequency of your waveform. Note that the
multimeter outputs RMS voltage for an AC signal.
3.
Set the oscilloscope channel 1
COUPLING
to
AC
. Press the yellow
button ‘1’. Go to
‘Coupling DC’ menu and use the ‘Multipurpose’ knob to select ‘AC’.
Adjust the
Vertical
and Horizontal Scale
knobs to have at least one complete cycle.
4.
Observe the signal with the digital multimeter and oscilloscope. Record the V
pp
, V
p
, V
rms
,
and period of the signal.
5.
Adjust the amplitude on the function generator to output a 5.0 V
rms
, 100 Hz. sinewave.
Record the V
pp
, V
p
, V
rms
, and period.
6.
Set the function generator output frequency to 10,000 Hz. Record the V
pp
, V
p
, V
rms
, and
period.
Assignment 3: Two Channel Analysis
In Assignment 2, the digital multimeter was used to quantify the amplitude of an AC signal. In
order to quantify a DC signal, the multimeter AC/DC switch must be set to DC. The oscilloscope
is capable of measuring two signals simultaneously. Each signal can be comprised of an AC and
DC component. If only the AC component of the signal is to be monitored, the oscilloscope
should be set to AC coupling. If both the AC and DC components of the signal are desired, the
oscilloscope should be set to DC coupling. The objective of this assignment is to introduce two
channel analysis using the oscilloscope and examine the effects of AC/DC coupling.
Procedure
IMPORTANT:
Whenever you generate a new signal, use the multimeter and oscilloscope to
check the magnitude and frequency of your waveform. You may need to disconnect your
cables from the oscilloscope in order to use the multimeter.
1.
Figure 2 shows the proper equipment configuration. Adjust the cabling according to the
new block diagram given below.
The output from the function generator should remain connected to
CH. 1
of the
oscilloscope. At this time, connect the output from the power supply to
CH. 2
of the
oscilloscope.
2.
Turn on
CH. 2
by pressing the blue button ‘2’ if this is off
. This will allow channels 1
and 2 to be viewed simultaneously.
Your preview ends here
Eager to read complete document? Join bartleby learn and gain access to the full version
- Access to all documents
- Unlimited textbook solutions
- 24/7 expert homework help
Department of Mechanical Engineering
MECH.3020
University of Massachusetts Lowell
Laboratory 2
MECH. 3020
P a g e
4 | 6
DC Power
Supply
Multimeter
Function
Generator
CH. 1
CH. 2
Oscilloscope
Figure 2. Block diagram for Assignment 3
3.
Set the output of the function generator to 5 V
rms
at 1000 Hz.
4.
Adjust the
Vertical Scale
of
CH. 2
to
5 Volts/DIV
.
5.
Set the oscilloscope channel 2
COUPLING
to
DC
.
Press the blue button ‘2’. Go to
‘Coupling DC’ menu and use the ‘Multipurpose’ knob to select ‘DC’ if it was not
selected before.
6.
Adjust the output voltage from the power supply to 10 Volts.
7.
Record the V
pp
, V
p
, V
rms
and periods associated with channels 1 and 2.
8.
Set the oscilloscope channel 2
COUPLING
to
AC
. Where did your DC signal go?
Change the coupling back to
DC
.
Department of Mechanical Engineering
MECH.3020
University of Massachusetts Lowell
Laboratory 2
MECH. 3020
P a g e
5 | 6
Assignment 4:
Digital Multimeter Upper Usable Frequency Range
The multimeter can be used to measure the RMS voltage of an AC signal, but it has an upper
usable frequency called the 3dB down point. The 3 dB down point is defined as the frequency at
which the measured voltage (V
o
) drops to 1/2
0.5
of the actual input voltage (V
i
). The 3 dB down
point will be determined by generating a sinusoidal wave at a constant amplitude with the
function generator. The amplitude of the signal will be recorded with the multimeter and
oscilloscope at various frequencies. The 3dB down point can then be determined by plotting the
ratio of the measured voltage (V
o
) to the input voltage (V
i
) in dB versus log frequency. The
frequency at which the ratio drops by 3 dB is the 3dB down point (also called the cutoff
frequency).
in
out
10
10
dB
V
V
log
20
G
log
20
G
=
=
Procedure
1.
Equipment configuration will be identical to Assignment 2 (refer to Figure 1).
2.
Set the output from the function generator to 5 volts peak at 100 Hz. Make sure the
AC
coupling
for
CH. 1
has been selected.
3.
Vary the output frequency from 100 Hz. to approximately 1 MHz in 10 to 15 steps. At
each frequency setting, record the voltage on the oscilloscope (V
i
) and on the digital
multimeter (V
o
).
4.
Plot the dB value versus frequency (logarithmic scale for x-axis). Determine where the
multimeter begins to become inaccurate. Increase the resolution of the measurement as
needed.
Assignment 5:
Analog Filter Effects on Signals of Various Frequencies
An analog filter can be used to limit the effects of signals outside the range of interest by limiting
the amplitude of these unwanted signals.
Different filters have different pass, transition, and
stop-band characteristics.
The characteristics of the 4-pole Butterworth filter will be explored.
Procedure
1.
Equipment configuration will be the same as in Assignment 4 except that the filter must
be wired in series between the function generator and multimeter.
2.
Turn on the filter and make sure “channel” is set to 1.
Type in 2000 and press “freq.”
The screen under “cutoff frequency” should read 2.000 and the “freq” and “kilo” buttons
should be lit.
(Note:
2 kHz was set as the cutoff frequency for the filter because it is
known from assignment #4 that the 3 dB down point of the multimeter is much higher
than 2 kHz.
This will limit the effects of the roll-off in the measurements caused by the
multimeter.)
Department of Mechanical Engineering
MECH.3020
University of Massachusetts Lowell
Laboratory 2
MECH. 3020
P a g e
6 | 6
3.
Vary the output frequency from 100 Hz. to approximately 5 kHz. in 10 to 15 steps
(should still be a 5 volt peak sinusoid). At each frequency setting, record the voltage on
the oscilloscope (V
i
) and on the digital multimeter (V
o
).
4.
Plot the dB value versus frequency (logarithmic scale for x-axis).
Increase the resolution
of the measurement as needed. This plot is representative of the characteristics of the
filter used.
Post-Lab Analysis
NOTE:
Be brief, concise and to the point in all responses.
Provide clear, concise answers to
questions.
Lengthy responses that ramble will not be graded and may lose additional points.
Q1.
Tabulate the quantities measured in Assignments 2 and 3. Compare the quantities
measured by both the oscilloscope and multimeter. (15 pt.)
Q2.
What is the primary difference between the oscilloscope and digital multimeter? (10 pt.)
Q3.
Explain the difference between AC and DC coupling. (10 pt.)
Q4.
What purpose does the function generator serve? (10 pt.)
Q5.
What is the 3 dB down point of the Fluke multimeter based on the measurements? (15
pt.)
Q6.
Make a proper engineering plot of the V
o
/V
i
in dB vs. frequency from Assignment 4. Use
logarithmic scale for x-axis. (15 pt.)
Q7.
Make a proper engineering plot the V
o
/V
i
in dB vs. frequency from Assignment 5. Use
logarithmic scale for x-axis. (15 pt.)
Q8.
Two sinusoidal signals are known to exist within a given signal, a 5 V
rms
100 Hz signal
and a 2 V
rms
3 kHz signal, but only the 100 Hz signal is of interest. If the analog filter
from Assignment 5 is used with the same 2 kHz cutoff frequency, what will be the
amplitude of both portions of the signal in voltage after filtering? (Hint: Use the plot from
Question 7) Are the effects of the 3 kHz signal ever completely removed? (10 pt.)
Your preview ends here
Eager to read complete document? Join bartleby learn and gain access to the full version
- Access to all documents
- Unlimited textbook solutions
- 24/7 expert homework help
Related Documents
Related Questions
b) A linear variable displacement transducer (LVDT) senses displacement and indicates a
voltage output, which is linear to the input. Figure 1 shows an LVDT setup used for static
calibration. It uses a micrometre to apply the known displacement and a voltmeter for the
output.
i) Suggest the independent, dependent and extraneous variables in this calibration.
ii) Indicate the basic functional elements of the device in Figure 1 in the form of block
diagram.
10
Micrometer
Fixed mounts
Connector
Excitation
voltage
in
Figure 1
Primary coil
xx_w_
LVDT
E
Output
voltage
10.135
Voltmeter
Secondary coil
Core
arrow_forward
Truncation errors are increased as the round-off errors are decreased.Group of answer choices True False
Say, you have a thermometer and you are checking the temperature of a body that has a temperature of 36o Using your thermometer five times, it gives you the following measurements: 29oC, 29.2oC, 29.3oC, 28.9oC, and 29.1oC. What can we conclude about the accuracy and the precision of the thermometer?Group of answer choices The thermometer is not accurate and not precise The thermometer is faulty. The thermometer is accurate and precise The thermometer is not accurate but precise.
Say, you have a thermometer and you are checking the temperature of a body that has a temperature of 36o Using your thermometer five times, it gives you the following measurements: 36oC, 35.6oC, 36oC, 37oC, and 36.2oC. What can we conclude about the accuracy and the precision of the thermometer?Group of answer choices The thermometer is accurate and precise. The thermometer is accurate but not precise. The…
arrow_forward
Subject: Mechanical Measurements
Do not copy online solutions. It's different value
arrow_forward
Pressurized eyes Our eyes need a certain amount of internal pressure in order to work properly, with the normal range being between 10 and 20 mm of mercury. The pressure is determined by a balance between the fluid entering and leaving the eye. If the pressure is above the normal level, damage may occur to the optic nerve where it leaves the eye, leading to a loss of the visual field termed glaucoma. Measurement of the pressure within the eye can be done by several different noninvasive types of instruments, all of which measure the slight deformation of the eyeball when a force is put on it. Some methods use a physical probe that makes contact with the front of the eye, applies a known force, and measures the deformation. One non-contact method uses a calibrated “puff” of air that is blown against the eye. The stagnation pressure resulting from the air blowing against the eyeball causes a slight deformation, the magnitude of which is correlated with the pressure within the eyeball.…
arrow_forward
4:07
3G l.
Moodle
A bbb2.du.edu.om
EE
...Session (15-3-
Said Grami O
College of Engineering
Vision for the Future
Question 3
During a step function calibration, a first-order instrument
is exposed to a step change of 200 units.
1- If after 1.5 s the instrument indicates 140 units, find the
instrument time constant.
2- find the error in the indicated value after 2 s
S.
3- Find the raise time at 90 % of the system.
y(0)=0 units; K=1 unit/unit.
Question 4
A first order instrument with a time constant of 2 seconds is to
be used to measure a periodic input. If a dynamic error of
±2% can be tolerated,
1- determine the maximum frequency of periodic input that
can be measured.
2- What is the associated time lag (in seconds) at that
frequency?
MECH374: Instrumentation and Measurement
Said Grami
Practice
arrow_forward
Chapter 12 - Lecture Notes.pptx: (MAE 272-01) (SP25) DY...
Scores
arrow_forward
Stiffness is the property that enables a material to withstand high stress without great strain. It is a resistance to any sort of deformation and is a function of the modulus of elasticity, E, of the material.
True
False
There are currently two (2) systems of units used in engineering practice in the United States; they are the U.S. Customary System, also referred to as the English System, and the International System or the SI System.
True
False
(explantion is not needed, just a correct answer, very simple, thank you)
arrow_forward
Biomechanies
arrow_forward
Instrumentation & Measurements
This homework measures your capability to design/analyze various components/variables of ameasurement system based on what you have studied.
Question is Attached in image. Thank you.
arrow_forward
7
arrow_forward
APPLICATIONS OF FIRST ORDER DIFFERENTIAL EQUATIONS
Newton’s Law of Cooling and Joint Proportions
At 4:00pm, a thermometer reading of 28 deg C is taken outside where the ambient temperature is –11 deg C. At 4:05 pm, the thermometer reads 8 deg above zero. After a while, the thermometer is returned to the room maintained at 28 deg C. At 4:12pm, the thermometer reads 15 deg C. When was the thermometer returned to the room?
arrow_forward
Basic Manufacturing Process with 2 Job Types +
Inspection
Time between job arrivals at a machining station is exponentially distributed with mean 4.4 minutes.
There are 2 types of jobs to be processed 30% of which is Type 1 and, 70% are of Type 2. Processing
times are exponentially distributed. Mean processing time for Type 1 is 4.8 minutes, for Type 2 it is
2.5 minutes.
After the job is processed, they go through an inspection process with one single inspector and an
inspection time with triangular distribution (1,2,3.5). Inspector decides whether the part is good
enough, scrap or should be reworked. 80% of the parts produced is good, 10 % is scrap and the rest
needs rework.
Rework is done by the same manufacturing machine. The priority among the parts will be Part1 first,
part2 second and reworks of both type comes later. Rework time is normally distributed with mean 2
minutes and 0,2 std dev.
Simulate the system for one 8-hour day.
arrow_forward
Chapter 3:
13. What type of geometric tolerance has no datum features?
14. A theoretically exact dimension is called a
15. The
feature symbol specifies that a group of two or more interrupted features of size are to be considered one single feature of size.
16. For an individual regular feature of size, no element of the feature shall extend beyond the maximum material condition boundary (envelope) of perfect form. This statement is the essence of
17. What is the name of the symbol used to identify physical features of a part as a datum feature and must not be applied to centerlines, center planes, or axes?
18. The
is the condition in which a feature of size contains the maximum amount of material within the stated limits of size.
19. The total positional tolerance equals the sum of the
tolerance and the
tolerance.
arrow_forward
I need help solving this problem.
arrow_forward
Hartley Electronics, Inc., in Nashville, producesshort runs of custom airwave scanners for the defense industry.The owner, Janet Hartley, has asked you to reduce inventory byintroducing a kanban system. After several hours of analysis, youdevelop the following data for scanner connectors used in onework cell. How many kanbans do you need for this connector?Daily demand 1,000 connectorsLead time 2 daysSafety stock 12 dayKanban size 500 connectors
arrow_forward
You’re testing a system that involves a fan, water pump, and compressor. The components are turned ON in sequence and sensor readings are taken, as per a written protocol that you’re following. The system is running as expected after the fan and water pump are turned ON. However, when you turn ON the compressor, it runs for about thirty seconds then stops, consistently.
Swapping out the compressor for a different one seems to solve the problem. What could have been the issue with the bad compressor?
arrow_forward
Hello, so i have attached two images. the first image is the quetions, i need part b) answered if possible. i have attached my findings to part a) to add to the little information we know to help with part b if needed. Thnks
arrow_forward
Need help filling out the table with the data I have (1st row is fine but show work please)
arrow_forward
Identify the measurement system stages for the following systems (refer back to Figure 1.5 and useother resources, such as a library or Internet search, as needed to learn more about each system):
d. Antilock braking system (automobile)e. Automobile cruise controlf. Throttle position sensor for a Formula One race car
arrow_forward
SEE MORE QUESTIONS
Recommended textbooks for you

Elements Of Electromagnetics
Mechanical Engineering
ISBN:9780190698614
Author:Sadiku, Matthew N. O.
Publisher:Oxford University Press

Mechanics of Materials (10th Edition)
Mechanical Engineering
ISBN:9780134319650
Author:Russell C. Hibbeler
Publisher:PEARSON

Thermodynamics: An Engineering Approach
Mechanical Engineering
ISBN:9781259822674
Author:Yunus A. Cengel Dr., Michael A. Boles
Publisher:McGraw-Hill Education

Control Systems Engineering
Mechanical Engineering
ISBN:9781118170519
Author:Norman S. Nise
Publisher:WILEY

Mechanics of Materials (MindTap Course List)
Mechanical Engineering
ISBN:9781337093347
Author:Barry J. Goodno, James M. Gere
Publisher:Cengage Learning

Engineering Mechanics: Statics
Mechanical Engineering
ISBN:9781118807330
Author:James L. Meriam, L. G. Kraige, J. N. Bolton
Publisher:WILEY
Related Questions
- b) A linear variable displacement transducer (LVDT) senses displacement and indicates a voltage output, which is linear to the input. Figure 1 shows an LVDT setup used for static calibration. It uses a micrometre to apply the known displacement and a voltmeter for the output. i) Suggest the independent, dependent and extraneous variables in this calibration. ii) Indicate the basic functional elements of the device in Figure 1 in the form of block diagram. 10 Micrometer Fixed mounts Connector Excitation voltage in Figure 1 Primary coil xx_w_ LVDT E Output voltage 10.135 Voltmeter Secondary coil Corearrow_forwardTruncation errors are increased as the round-off errors are decreased.Group of answer choices True False Say, you have a thermometer and you are checking the temperature of a body that has a temperature of 36o Using your thermometer five times, it gives you the following measurements: 29oC, 29.2oC, 29.3oC, 28.9oC, and 29.1oC. What can we conclude about the accuracy and the precision of the thermometer?Group of answer choices The thermometer is not accurate and not precise The thermometer is faulty. The thermometer is accurate and precise The thermometer is not accurate but precise. Say, you have a thermometer and you are checking the temperature of a body that has a temperature of 36o Using your thermometer five times, it gives you the following measurements: 36oC, 35.6oC, 36oC, 37oC, and 36.2oC. What can we conclude about the accuracy and the precision of the thermometer?Group of answer choices The thermometer is accurate and precise. The thermometer is accurate but not precise. The…arrow_forwardSubject: Mechanical Measurements Do not copy online solutions. It's different valuearrow_forward
- Pressurized eyes Our eyes need a certain amount of internal pressure in order to work properly, with the normal range being between 10 and 20 mm of mercury. The pressure is determined by a balance between the fluid entering and leaving the eye. If the pressure is above the normal level, damage may occur to the optic nerve where it leaves the eye, leading to a loss of the visual field termed glaucoma. Measurement of the pressure within the eye can be done by several different noninvasive types of instruments, all of which measure the slight deformation of the eyeball when a force is put on it. Some methods use a physical probe that makes contact with the front of the eye, applies a known force, and measures the deformation. One non-contact method uses a calibrated “puff” of air that is blown against the eye. The stagnation pressure resulting from the air blowing against the eyeball causes a slight deformation, the magnitude of which is correlated with the pressure within the eyeball.…arrow_forward4:07 3G l. Moodle A bbb2.du.edu.om EE ...Session (15-3- Said Grami O College of Engineering Vision for the Future Question 3 During a step function calibration, a first-order instrument is exposed to a step change of 200 units. 1- If after 1.5 s the instrument indicates 140 units, find the instrument time constant. 2- find the error in the indicated value after 2 s S. 3- Find the raise time at 90 % of the system. y(0)=0 units; K=1 unit/unit. Question 4 A first order instrument with a time constant of 2 seconds is to be used to measure a periodic input. If a dynamic error of ±2% can be tolerated, 1- determine the maximum frequency of periodic input that can be measured. 2- What is the associated time lag (in seconds) at that frequency? MECH374: Instrumentation and Measurement Said Grami Practicearrow_forwardChapter 12 - Lecture Notes.pptx: (MAE 272-01) (SP25) DY... Scoresarrow_forward
- Stiffness is the property that enables a material to withstand high stress without great strain. It is a resistance to any sort of deformation and is a function of the modulus of elasticity, E, of the material. True False There are currently two (2) systems of units used in engineering practice in the United States; they are the U.S. Customary System, also referred to as the English System, and the International System or the SI System. True False (explantion is not needed, just a correct answer, very simple, thank you)arrow_forwardBiomechaniesarrow_forwardInstrumentation & Measurements This homework measures your capability to design/analyze various components/variables of ameasurement system based on what you have studied. Question is Attached in image. Thank you.arrow_forward
- 7arrow_forwardAPPLICATIONS OF FIRST ORDER DIFFERENTIAL EQUATIONS Newton’s Law of Cooling and Joint Proportions At 4:00pm, a thermometer reading of 28 deg C is taken outside where the ambient temperature is –11 deg C. At 4:05 pm, the thermometer reads 8 deg above zero. After a while, the thermometer is returned to the room maintained at 28 deg C. At 4:12pm, the thermometer reads 15 deg C. When was the thermometer returned to the room?arrow_forwardBasic Manufacturing Process with 2 Job Types + Inspection Time between job arrivals at a machining station is exponentially distributed with mean 4.4 minutes. There are 2 types of jobs to be processed 30% of which is Type 1 and, 70% are of Type 2. Processing times are exponentially distributed. Mean processing time for Type 1 is 4.8 minutes, for Type 2 it is 2.5 minutes. After the job is processed, they go through an inspection process with one single inspector and an inspection time with triangular distribution (1,2,3.5). Inspector decides whether the part is good enough, scrap or should be reworked. 80% of the parts produced is good, 10 % is scrap and the rest needs rework. Rework is done by the same manufacturing machine. The priority among the parts will be Part1 first, part2 second and reworks of both type comes later. Rework time is normally distributed with mean 2 minutes and 0,2 std dev. Simulate the system for one 8-hour day.arrow_forward
arrow_back_ios
SEE MORE QUESTIONS
arrow_forward_ios
Recommended textbooks for you
- Elements Of ElectromagneticsMechanical EngineeringISBN:9780190698614Author:Sadiku, Matthew N. O.Publisher:Oxford University PressMechanics of Materials (10th Edition)Mechanical EngineeringISBN:9780134319650Author:Russell C. HibbelerPublisher:PEARSONThermodynamics: An Engineering ApproachMechanical EngineeringISBN:9781259822674Author:Yunus A. Cengel Dr., Michael A. BolesPublisher:McGraw-Hill Education
- Control Systems EngineeringMechanical EngineeringISBN:9781118170519Author:Norman S. NisePublisher:WILEYMechanics of Materials (MindTap Course List)Mechanical EngineeringISBN:9781337093347Author:Barry J. Goodno, James M. GerePublisher:Cengage LearningEngineering Mechanics: StaticsMechanical EngineeringISBN:9781118807330Author:James L. Meriam, L. G. Kraige, J. N. BoltonPublisher:WILEY

Elements Of Electromagnetics
Mechanical Engineering
ISBN:9780190698614
Author:Sadiku, Matthew N. O.
Publisher:Oxford University Press

Mechanics of Materials (10th Edition)
Mechanical Engineering
ISBN:9780134319650
Author:Russell C. Hibbeler
Publisher:PEARSON

Thermodynamics: An Engineering Approach
Mechanical Engineering
ISBN:9781259822674
Author:Yunus A. Cengel Dr., Michael A. Boles
Publisher:McGraw-Hill Education

Control Systems Engineering
Mechanical Engineering
ISBN:9781118170519
Author:Norman S. Nise
Publisher:WILEY

Mechanics of Materials (MindTap Course List)
Mechanical Engineering
ISBN:9781337093347
Author:Barry J. Goodno, James M. Gere
Publisher:Cengage Learning

Engineering Mechanics: Statics
Mechanical Engineering
ISBN:9781118807330
Author:James L. Meriam, L. G. Kraige, J. N. Bolton
Publisher:WILEY