(a)
Interpretation:
The amount by which
Concept introduction:
Limiting condition of operational amplifier as a comparator implies that the comparator output voltage reaches the limiting value of output which is given here as +13 V on the positive side and
We have,
Corresponding to the positive saturation value and negative saturation value respectively. We need to find the difference in input voltage in each case.
Answer to Problem 3.1QAP
Explanation of Solution
Given the open loop gain A is 100,000
Operational amplifier as a comparator compares two voltages at its respective input terminals
We have to find the amount by which
Case 1: For
Here,
Hence,
This implies
Case 2: For
Here,
Hence,
This implies
(b)
Interpretation:
The amount by which
Concept introduction:
Limiting condition of operational amplifier as a comparator implies that the comparator output voltage reaches the limiting value of output which is given here as +13 V on the positive side and
We have,
Corresponding to the positive saturation value and negative saturation value respectively. We need to find the difference in input voltage in each case.
Answer to Problem 3.1QAP
Explanation of Solution
Given the open loop gain A is 600,000.
Operational amplifier as a comparator compares two voltages at its respective input terminals
We have to find the amount by which
Case 1: For
Here,
Hence,
This implies
Case 2: For
Here,
Hence,
This implies
(c)
Interpretation:
The amount by which
Concept introduction:
Limiting condition of operational amplifier as a comparator implies that the comparator output voltage reaches the limiting value of output which is given here as +13 V on the positive side and
We have,
Corresponding to the positive saturation value and negative saturation value respectively. We need to find the difference in input voltage in each case.
Answer to Problem 3.1QAP
Explanation of Solution
Given the open loop gain A is
Operational amplifier as a comparator compares two voltages at its respective input terminals
We have to find the amount by which
Case 1: For
Here,
Hence,
This implies
Case 2 : For
Here,
Hence,
This implies
Want to see more full solutions like this?
Chapter 3 Solutions
Principles of Instrumental Analysis
- Kindly answer no iii, iv & varrow_forwardP1B.1 A rotating slotted-disc apparatus consists of five coaxial 5.0 cm diameter discs separated by 1.0 cm, the radial slots being displaced by 2.0° between neighbours. The relative intensities, I, of the detected beam of Kr atoms for two different temperatures and at a series of rotation rates were as follows: v/Hz 20 40 80 100 120 I (40 K) 0.846 0.513 0.069 0.015 0.002 I (100 K) 0.592 0.485 0.217 0.119 0.057 Find the distributions of molecular velocities, f(v,), at these temperatures, and check that they conform to the theoretical prediction for a one-dimensional system for this low-pressure, collision-free system.arrow_forwardKindly answer no i, ii & iiiarrow_forward
- 3) The solution taken from a factory waste sample to prepare an internal standard contains Pb + 2 and Ag + 2, and the analysis peaks were determined as 298 mAU and 324 mAU. To analyze the unknown, a solution containing 0.150 M, 50.0 mL of Ag + 2 is put into the unknown liquid and completed to 100.0 mL. When the peaks are measured again, 670 mAU and 856 mAU signals are read from the samples. What is the concentration of Pb + 2 in the unknown liquid?arrow_forwardThe molar absorptivity for aqueous solutions of phenol at 211 nm is 6.17 x 103 L cml mol-'. Calculate the permissible range of phenol concentrations if the transmittance is to be less than 85% and greater than 7% when the measurements are made in 1.00-cm cells.arrow_forwardThe student was supposed to measure the pH of a 0.1 M hydrochloric acid solution. For this purpose, he used a pH-meter and obtained the following results: 0.90, 0.93, 1.04, 1.02. Present the appropriate graphic and statistical elaboration of the performed analysis and confirm whether the applied pH-meter worked properly.arrow_forward
- A standard solution was prepared containing 10.0 ppm of an analyte and 15.0 ppm of an internal standard. Analysis of the sample gave signals for the analyte and internal standard of 0.155 and 0.233 (arbitrary units), respectively. Sufficient internal standard was added to an unknown sample to make it 15.0 ppm in the internal standard. Analysis of the unknown sample yielded signals for the analyte and internal standard of 0.274 and 0.198, respectively. Calculate the concentration of analyte in the unknown sample.arrow_forward2. A gas at 300 K and 1 bar (state 1) flows through an adiabatic compressor and exits at 20 bar and 700 K (state 2). The gas obeys the following equation of state, where a and B are constants. v-B The values are given as a = 10-3 _m³k bar mol the gas at 1 bar is Cp = 30 mol K RT ар + P T and B = 8 x 10-5 m³. The heat capacity of molarrow_forwardGiven the equation of a calibration curve y=5.63x + 0.006, identify the variable that equates to the absorbance value of the Beer's law equation.arrow_forward
- (B) A 0.05 mM solution of DCPIP was placed in a 1 cm spectrophotometer cuvette and a beam of light at 600nm was passed through the cuvette. The extinction coefficient for DCPIP is 3500 L/mole/cm under the specific conditions used. What proportion (as a percentage) of light would emerge from the other side of the cuvette?arrow_forwardGiven these two equations, calculate Cv for CO2 and N2 using the velocity of sound. 6. According to statistical mechanics, the contribution to Cy due to the ith vibrational mode of a molecule is c ee;/T Cv,vib,4 = R(7) ³ (e®,/T — 1)² › where Oi = hcv./kB, h is Planck's constant, and c is the speed of light in cm s-¹. v; is the vibrational frequency of the ith vibrational mode in wavenumbers (cm-¹). The quantity O, has dimensions of temperature and is sometimes referred to as the vibrational tem- perature (of the ith vibrational mode). The total vibrational contribution to the molar heat capacity for an N-atom (linear) molecule is then Cy,vib Cv,vib,i Calculate Cy for CO₂ at 300 K, assuming that the translational and rotational con- tributions are (3/2)R and R, respectively, and use the preceding equations to obtain the vibrational contribution. Four vibrational modes for CO₂ are expected (3N- 5). How- ever, three fundamental frequencies are observed at v= 667.3, 1340, and 2349.3 cm-¹.…arrow_forwardDesign a voltage divider biased common emitter amplifier with a bypassed emitter resistance for a DC supply of Vcc D (18+R)V and a quiescent current of 2mA. The voltage drop across collector should be 40% of DC supply voltage For the given design conditions: R = 02 Compute the circuit resistances.biased values of voltages and currents. Draw the completely labelled circuit diagram.arrow_forward
- Principles of Instrumental AnalysisChemistryISBN:9781305577213Author:Douglas A. Skoog, F. James Holler, Stanley R. CrouchPublisher:Cengage Learning