Chapter 15, Problem 15.1QAP
Interpretation Introduction

Interpretation:

The points based on which a fluorescence emission spectrum and a fluorescence excitation spectrum can be differentiated needs to be explained. From the two spectrums, the one that resembles an absorption spectrum more closely is to be stated.

Concept introduction:

Fluorescence is a word used for the members of the luminescence family. In this family, the molecules of a substance can emit their own light with the help of electronically excited states which can be formed by physical, chemical or mechanical mechanism.

Expert Solution & Answer
Check Mark

Explanation of Solution

There are two fluorescence spectrums namely the fluorescence emission spectrum and the fluorescence excitation spectrum. The difference between the two is shown below.

Fluorescence Emission Spectrum Fluorescence Excitation Spectrum
The wavelength for excitation does not vary. It is constant. The wavelength for excitation varies. It is not constant.
The intensity of emission changes according to the change in emission wavelength. The intensity of emission does not change. It is measured at one wavelength.

In the fluorescence excitation spectrum, the emission intensity is proportional to the absorbance of the molecule. This phenomenon makes it closer to absorption spectrum rather thanfluorescence emission spectrum.

Thus, the fluorescence excitation spectrum is closer to an absorption spectrum.

Want to see more full solutions like this?

Subscribe now to access step-by-step solutions to millions of textbook problems written by subject matter experts!
Students have asked these similar questions
What is the effect of temperature on atomic spectroscopy?
White light passes through a white or clear sample and is collected by a detector. Which absorption spectra could be correct? Select all possible answers. Absorbance O B -400 nm Absorbance -700 nm B -400 nm wavelength A Absorbance wavelength 400 nm 700 nm wavelength
Substances A and B have retention times of 15.32 min and 16.92 min, respectively, on a 50.0-cm column. An unretained species passes through the column in 1.15 min. The peak widths (at base) for A and B are 1.11 min and 1.21 min, respectively. What is the time required to elute substance B on the column that gives a.resolution of 1.8? 16R, H () (tr)B 2 (1+kB) (kB)² %3D
Knowledge Booster
Background pattern image
Chemistry
Learn more about
Need a deep-dive on the concept behind this application? Look no further. Learn more about this topic, chemistry and related others by exploring similar questions and additional content below.
Similar questions
SEE MORE QUESTIONS
Recommended textbooks for you
Text book image
Principles of Instrumental Analysis
Chemistry
ISBN:9781305577213
Author:Douglas A. Skoog, F. James Holler, Stanley R. Crouch
Publisher:Cengage Learning
Text book image
Chemistry: Matter and Change
Chemistry
ISBN:9780078746376
Author:Dinah Zike, Laurel Dingrando, Nicholas Hainen, Cheryl Wistrom
Publisher:Glencoe/McGraw-Hill School Pub Co
Text book image
Fundamentals Of Analytical Chemistry
Chemistry
ISBN:9781285640686
Author:Skoog
Publisher:Cengage
Text book image
Physical Chemistry
Chemistry
ISBN:9781133958437
Author:Ball, David W. (david Warren), BAER, Tomas
Publisher:Wadsworth Cengage Learning,
Text book image
Chemistry for Engineering Students
Chemistry
ISBN:9781337398909
Author:Lawrence S. Brown, Tom Holme
Publisher:Cengage Learning
Text book image
Chemistry for Engineering Students
Chemistry
ISBN:9781285199023
Author:Lawrence S. Brown, Tom Holme
Publisher:Cengage Learning