You develop a sensor with a thermoelectric cooled CCD detector (-20C). You get a sensitivity of 15000 counts/molarity and a method detection limit 0.1 mM (millimolar). Assuming all the noise is Dark Noise (slide 55), what would be the sensitivity and detection limit if the detector were only cooled to 10C?

icon
Related questions
Question
You develop a sensor with a thermoelectric cooled CCD
detector (-20C). You get a sensitivity of 15000
counts/molarity and a method detection limit 0.1 mM
(millimolar). Assuming all the noise is Dark Noise (slide
55), what would be the sensitivity and detection limit if the
detector were only cooled to 10C?
Transcribed Image Text:You develop a sensor with a thermoelectric cooled CCD detector (-20C). You get a sensitivity of 15000 counts/molarity and a method detection limit 0.1 mM (millimolar). Assuming all the noise is Dark Noise (slide 55), what would be the sensitivity and detection limit if the detector were only cooled to 10C?