You are asked to design a spectral filter that practically removes 99.0% of the low energy photons in an X-ray beam. Such photons contribute to the patient dose without contributing to the image and are defined as no more than 1% of these photons making it to the other side of the patient. Assume a patient can be modelled as a 20cm thick homogenous object with linear attenuation coefficients as shown below in Table 1. What is the thickness of the filter needed to eliminate all the energies which satisfy the above requirement? Filter linear attenuation properties are given below in Table 2. Table 1: Lincar attenuation coefficients vs. energy of patient equivalent material Energy [keV] Hjet (mm 20 0.02601 30 0.02407 40 0.02303 50 0.02151 60 0.02013 Table 2: Linear attenuation coefficients vs. energy of filter material Energy (keV Hr mm 20 30 40 50 60 70 0.1225 0.1067 0.0872 0.07541 0.06684 0.06327

icon
Related questions
Question
You are asked to design a spectral filter that practically removes 99.0% of the low energy photons in an X-ray beam. Such photons
contribute to the patient dose without contributing to the image and are defined as no more than 1% of these photons making it to the
other side of the patient. Assume a patient can be modelled as a 20cm thick homogenous object with linear attenuation coefficients as
shown below in Table 1. What is the thickness of the filter needed to eliminate all the energies which satisfy the above requirement?
Filter linear attenuation properties are given below in Table 2.
Table 1: Linear attenuation coefficients vs. energy of patient equivalent material
Energy [keV]
Habject (mm|
20
0.02601
30
0.02407
40
0.02303
50
60
0.02151
0.02013
Table 2: Linear attenuation coefficients vs. energy of filter material
60
0.06684
20
Energy (keV]
Hrter (mm
30
40
50
70
0.1225
0.1067
0.0872
0.07541
0.06327
Transcribed Image Text:You are asked to design a spectral filter that practically removes 99.0% of the low energy photons in an X-ray beam. Such photons contribute to the patient dose without contributing to the image and are defined as no more than 1% of these photons making it to the other side of the patient. Assume a patient can be modelled as a 20cm thick homogenous object with linear attenuation coefficients as shown below in Table 1. What is the thickness of the filter needed to eliminate all the energies which satisfy the above requirement? Filter linear attenuation properties are given below in Table 2. Table 1: Linear attenuation coefficients vs. energy of patient equivalent material Energy [keV] Habject (mm| 20 0.02601 30 0.02407 40 0.02303 50 60 0.02151 0.02013 Table 2: Linear attenuation coefficients vs. energy of filter material 60 0.06684 20 Energy (keV] Hrter (mm 30 40 50 70 0.1225 0.1067 0.0872 0.07541 0.06327
Expert Solution
Step 1

The filter asked to be designed should be able to remove 99% of the photons incident on it, and only let 1% of the photos pass through.

This can be done as the X-ray beam passing through the filter gets attenuated as it goes through.

The attenuated intensity is given in terms of the original intensity as

I=Ioe-μxIo is the original intensityμ is the attenuation factorx is the thickness of the material

Since 99% of the incident radiation is to be blocked, and only 1% is to be passed,

I=1%IoI=1100IoI=0.01Io

trending now

Trending now

This is a popular solution!

steps

Step by step

Solved in 3 steps

Blurred answer