Copper was determined in a river water sample by atomic absorption spectrometry and the method of standard additions. For the addition, 100.0 μL of a 1000.0-μg/mL Cu standard was added to 150.0 mL of solution. The following data were obtained: Absorbance of reagent blank = 0.009 Absorbance of sample = 0.454 Absorbance of sample plus addition – blank = 1.011 Calculate the copper concentration in the sample. Concentration = ____ μg/mL Later studies showed that the reagent blank used to obtain the above data was inadequate and that the actual blank absorbance was 0.060. Find the copper concentration with the appropriate blank, and determine the error caused by using an improper blank. Concentration = _____  μg/mL Error = _____ %

Fundamentals Of Analytical Chemistry
9th Edition
ISBN:9781285640686
Author:Skoog
Publisher:Skoog
Chapter28: Atomic Spectroscopy
Section: Chapter Questions
Problem 28.15QAP
icon
Related questions
Question
100%

Copper was determined in a river water sample by atomic absorption spectrometry and the method of standard additions. For the addition, 100.0 μL of a 1000.0-μg/mL Cu standard was added to 150.0 mL of solution. The following data were obtained:

Absorbance of reagent blank = 0.009

Absorbance of sample = 0.454

Absorbance of sample plus addition – blank = 1.011

  1. Calculate the copper concentration in the sample.

    Concentration = ____ μg/mL

  2. Later studies showed that the reagent blank used to obtain the above data was inadequate and that the actual blank absorbance was 0.060. Find the copper concentration with the appropriate blank, and determine the error caused by using an improper blank.

    Concentration = _____  μg/mL

    Error = _____ %