220 volts is supplied across 1200 winding of the primary coil of the autotransformer.If 1650 windings are tapped, what voltage will be supplied to the primary coil of thehigh-voltage transformer?2. A kVp meter reads 86 kVp and the turns ratio of the high-voltage step-up transformeris 1200. What is the true voltage across the meter?3. The supply voltage from the autotransformer to the filament transformer is 60 volts. If theturns ratio of the filament transformer is 1/12, what is the filament voltage?4. If the current in the primary side of the filament transformer in question 3 were 0.5 A,what would be the filament current?5. The supply to a high-voltage step-up transformer with a turns ratio of 550 is 190 volts.What is the voltage across the x-ray tube?
220 volts is supplied across 1200 winding of the primary coil of the autotransformer.
If 1650 windings are tapped, what voltage will be supplied to the primary coil of the
high-voltage transformer?
2. A kVp meter reads 86 kVp and the turns ratio of the high-voltage step-up transformer
is 1200. What is the true voltage across the meter?
3. The supply voltage from the autotransformer to the filament transformer is 60 volts. If the
turns ratio of the filament transformer is 1/12, what is the filament voltage?
4. If the current in the primary side of the filament transformer in question 3 were 0.5 A,
what would be the filament current?
5. The supply to a high-voltage step-up transformer with a turns ratio of 550 is 190 volts.
What is the voltage across the x-ray tube?