Old Measurements 010422
docx
keyboard_arrow_up
School
University of Central Arkansas *
*We aren’t endorsed by this school
Course
1400
Subject
Mechanical Engineering
Date
Feb 20, 2024
Type
docx
Pages
8
Uploaded by CountEnergyOryx35
Measurements 1
Measurements
v010422_PD
Objectives: The student will become familiar with some key pieces of glassware and instrumentation. The student will learn to record measurements correctly, assess the precision of a piece of glassware or instrument, and identify possible errors that can occur. Background: (Section E.2-E.4 in Tro 2
nd
Edition)
The quality of experimental measurements is affected by the number of
digits that can be reported for the equipment used for the measurement, the
number of measurements made, and the experimenter. When making a
measurement, the investigator should report all digits that can be read with
certainty plus one additional digit that is the estimated or uncertain digit. The
meaningful digits of a measurement are called significant figures.
Precision refers to the reproducibility of measurements. The closer the
measurements are to one another, the greater the precision. Increased precision
can be observed in measurements obtained from instruments that can be read to
a greater number decimal places with certainty. Accuracy refers to how close a measurement is to the real value. The
accuracy of the measurement reflects the quality of the instrument. An
instrument that can be read with more digits of certainty typically yields a more
accurate value assuming that the instrument has been properly calibrated. Errors occur during the collection of data. Errors are of two types: systematic and random. Systematic errors are caused by a defect in the analytical method, by an improperly functioning instrument, or by the analyst. Systematic errors cause a decrease in the accuracy of a set of measurements. Systematic errors can be corrected by calibration of instrumentation, alteration of the procedure, use of glassware and instruments that can be read to a greater number of significant figures, and by attentiveness of the analyst. Random errors cause a reduction in the precision of a series of measurements. Random errors are uncontrollable errors such as electrical fluctuations experienced by balances.
A single measurement is insufficient to determine the accuracy and precision of an instrument.
It is therefore common practice to make repetitive measurements for a value. The precision of a series of measurements is indicated using the mean of the results ± the error. Example: A student measured the mass of a tennis ball three times on a lab balance that read to
the first decimal place. They recorded the mass as 54.4 g, 54.7 g, and 54.1 g. The mean of all of these values is 54.4 g. Assuming reasonable balance calibration, it can be concluded that the true mass of the ball is the range of 54.1 g and 54.7 g. The student reports this more succinctly as
54.4 ± 0.3 g. The accuracy of a series of measurements can only be assessed if the true value is known. But, by using a calibrated instrument and making several measurements the probability that the mean of the measurements and the real value will be the same is improved.
https://youtu.be/sX_op1xBaHY
https://youtu.be/Y5sUB0suJ9A
Measurements 2
Safety:
Wear safety goggles. Wash your hands thoroughly before leaving the lab.
Procedure:
Measuring Lengths with Rulers (Rulers 1, 2, and 3 are printed on the next page)
1.
On your data sheet, indicate the number of digits past the decimal that a measurement made with Ruler 1 should contain.
2.
Use Ruler 1 to measure the length of the wooden block. Record you value, including units.
3.
Repeat steps 1 and 2 for Ruler 2 and Ruler 3.
4.
Report your data to the instructor.
5.
Once all measurements have been obtained the class results will be posted. Record the class results and determine the mean and ± error for each ruler.
Measuring Liquid Volumes using a Buret and Graduated Cylinder
1.
On your data sheet, indicate the number of decimal places that should be recorded for the buret and the number of decimal places that should be recorded for a 10-mL
graduated cylinder.
2.
Fill the buret to approximately the 0 mL mark with tap water. 3.
Read the initial water level of the buret (using the appropriate number of
digits). Record this under Trial 1. Include units.
4.
Dispense about 5 mL of tap water from the buret into a 10-mL graduated
cylinder.
5.
Record the new volume of the buret under Trial 1. Include units. 6.
Record the volume of the water in the graduated cylinder under Trial 1.
Include units.
7.
Empty and dry the graduated cylinder.
8.
Repeat steps 3-7 for Trial 2 and Trial 3 beginning with the current volume in the buret.
9.
Determine the difference between the buret and graduated cylinder readings.
Measuring Masses with Balances
1.
On your data sheet indicate the number of places past the decimal that should be recorded for the balance.
2.
Tare the balance.
3.
Place a rubber stopper on the balance and record the mass as mass 1, balance 1. Include units.
4.
Remove the object. Let the balance come to zero and tare balance again.
5.
Replace the stopper on the same balance and record the mass as mass 2, balance 1. Include units.
6.
Remove the object. Let the balance come to zero and tare balance for a third time.
7.
Replace the stopper on the same balance and record the mass as mass 3, balance 1. Include units.
8.
Repeat steps 2-7 two more times using two other balances but the same rubber stopper. The second balance used should be balance 2, and the third balance used should be balance 3.
https://youtu.be/t4K-3_H5kxU
F
Measurements 3
Data
and
Your preview ends here
Eager to read complete document? Join bartleby learn and gain access to the full version
- Access to all documents
- Unlimited textbook solutions
- 24/7 expert homework help
Measurements 4
Results
Measuring Lengths with Rulers
# of decimal places expected using Ruler 1
________
reading 1
st
ruler
________
# of decimal places expected using Ruler 2
________
reading 2
nd
ruler
________
# of decimal places expected using Ruler 3
________
reading 3
rd
ruler
________
Class Results
Ruler 1
Ruler 2
Ruler 3
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
Mean Measurement
Difference between Mean Measurement and Lowest Measurement
Difference between Mean Measurement and Highest Measurement
Mean ± Error
Ruler 1
________
________
________
_________________
Ruler 2
________
________
________
_________________
Ruler 3
________
________
________
_________________
Measurements 5
Measuring Liquid Volumes using a Buret and Graduated Cylinder
# of places past the decimal that should be listed from a buret
________
# of places past the decimal that should be listed from a 10-mL graduated cylinder
________
Trial 1
Trial 2
Trial 3
Initial Buret Reading
__________
__________
__________
Final Buret Reading
__________
__________
__________
Volume of Water Dispensed by Buret
__________
__________
__________
Volume of Water in Graduated Cylinder
__________
__________
__________
Difference between the Buret and
Graduated cylinder readings (abs. value)
__________
__________
__________
Measuring Masses with Balances
# of places past the that should be listed for a reading from the laboratory balance
________
Balance 1
Balance 2
Balance 3
Mass 1
____________
____________
____________
Mass 2
____________
____________
____________
Mass 3
____________
____________
____________
Mean Measurement
Difference between Mean Measurement and Lowest Measurement
Difference between Mean Measurement and Highest Measurement
Mean ± Error
Balance 1
________
________
________
_________________
Balance 2
________
________
________
_________________
Balance 3
________
________
________
_________________
Measurements 6
Pre-laboratory Assignment
1.
Errors in measurements can be classified as random or systematic. a.
Define random error.
b.
Define systematic error.
2.
A sample of salt was massed three times. The readings were 1.0639 g, 1.0636 g and 1.0642 g. Express the mass of the salt with the correct number of significant figures. Include the error of the measurement in your answer.
3.
A student measured the same block with two different rulers to obtain a length of 20.6 cm and 20.55 cm. Which of these measurements is more precise? Explain.
4.
A disreputable grocer places their thumb on the scale every time they weigh produce to sell to a customer. This introduces an obvious error in the measured mass of the produce. What kind: systematic or random?
Your preview ends here
Eager to read complete document? Join bartleby learn and gain access to the full version
- Access to all documents
- Unlimited textbook solutions
- 24/7 expert homework help
Measurements 7
Post-Laboratory Questions
1.
Which ruler, ruler 2 or ruler 3, from Part I exhibited the greatest amount of precision? Explain.
2.
What type of error, random or systematic, is responsible for the different readings obtained by the class for Ruler 3? Explain.
3.
(a) Which volume reading is more precise, that obtained from the buret or the 10-mL graduated cylinder? Explain.
(b) Compare the differences between the buret and graduated cylinder readings. Does your cylinder consistently read more than or less than the buret for the same volume?
(c) If you failed to properly dry the cylinder prior to dispensing the next trial from the buret, does this represent systematic or a random error?
4.
List 2 errors likely responsible for the fluctuation in readings obtained for mass of the stopper taken on balance 1. State whether these errors are random or systematic. Explain.
Measurements 8
5.
What error is likely responsible for the mass of the stopper changing depending on the balance
used? Is it a random error or systematic error? Explain.
6.
Identify each of the following as being due to random error or systematic error.
i.
A student was massing a block. As a classmate passed the balance the student noticed that the mass changed.
Random
Systematic
ii.
A student massed a block on one balance; he then massed it again on a different balance. The
mass changed.
Random
Systematic
iii.
Two students read the same buret. One read the volume of liquid as 10.91 mL, the other read the volume as 10.89 mL. Random
Systematic
Explain your answer to question 6 part III.
Related Documents
Related Questions
Truncation errors are increased as the round-off errors are decreased.Group of answer choices True False
Say, you have a thermometer and you are checking the temperature of a body that has a temperature of 36o Using your thermometer five times, it gives you the following measurements: 29oC, 29.2oC, 29.3oC, 28.9oC, and 29.1oC. What can we conclude about the accuracy and the precision of the thermometer?Group of answer choices The thermometer is not accurate and not precise The thermometer is faulty. The thermometer is accurate and precise The thermometer is not accurate but precise.
Say, you have a thermometer and you are checking the temperature of a body that has a temperature of 36o Using your thermometer five times, it gives you the following measurements: 36oC, 35.6oC, 36oC, 37oC, and 36.2oC. What can we conclude about the accuracy and the precision of the thermometer?Group of answer choices The thermometer is accurate and precise. The thermometer is accurate but not precise. The…
arrow_forward
A new electronic component for aircraft is tested to withstand temperatures as low as -50°F. One engineer suggests that the device cannot actually stand temperatures that low. What are the Type I and Type II errors?
Type I: The device CAN NOT withstand temperatures as low as -50F but the engineer incorrectly finds that it CAN.Type II: The device CAN withstand temperatures as low as -50F but the engineer incorrectly finds that it CAN NOT.
Type I: The device CAN withstand temperatures as low as -50F but the engineer incorrectly finds that it CAN NOT.Type II: The device CAN NOT withstand temperatures as low as -50F but the engineer incorrectly finds that it CAN.
Is the answer the first or second bullet point?
arrow_forward
Can someone please help me to answer all of the following questions thank you!!
arrow_forward
Question 4
arrow_forward
A. A student establishes the time constant of a temperature sensor by first holding it immersed in hot water and then
suddenly removing it and holding it immersed in cold water. Several other students perform the same test with
similar sensors. Overall, their results are inconsistent, with estimated time constants differing by as much as a
factor of 1.2. Offer suggestions about why this might happen. Hint: Try this yourself and think about control of
test conditions.
B. Which would you expect to be better suited to measure a time-dependent temperature, a thermal sensor (e.g., a
thermocouple) having a small diameter spherical bead or one having a large diameter spherical bead? Why?
arrow_forward
I need help solving this problem.
arrow_forward
Chapter 12 - Lecture Notes.pptx: (MAE 272-01) (SP25) DY...
Scores
arrow_forward
Question 2
You are a biomedical engineer working for a small orthopaedic firm that fabricates rectangular shaped fracture
fixation plates from titanium alloy (model = "Ti Fix-It") materials. A recent clinical report documents some problems with the plates
implanted into fractured limbs. Specifically, some plates have become permanently bent while patients are in rehab and doing partial
weight bearing activities.
Your boss asks you to review the technical report that was generated by the previous test engineer (whose job you now have!) and used to
verify the design. The brief report states the following... "Ti Fix-It plates were manufactured from Ti-6Al-4V (grade 5) and machined into
solid 150 mm long beams with a 4 mm thick and 15 mm wide cross section. Each Ti Fix-It plate was loaded in equilibrium in a 4-point bending
test (set-up configuration is provided in drawing below), with an applied load of 1000N. The maximum stress in this set-up was less than the
yield stress for the…
arrow_forward
b) A linear variable displacement transducer (LVDT) senses displacement and indicates a
voltage output, which is linear to the input. Figure 1 shows an LVDT setup used for static
calibration. It uses a micrometre to apply the known displacement and a voltmeter for the
output.
i) Suggest the independent, dependent and extraneous variables in this calibration.
ii) Indicate the basic functional elements of the device in Figure 1 in the form of block
diagram.
10
Micrometer
Fixed mounts
Connector
Excitation
voltage
in
Figure 1
Primary coil
xx_w_
LVDT
E
Output
voltage
10.135
Voltmeter
Secondary coil
Core
arrow_forward
Do not provide handwritten solution,Maintain accuracy and quality in your answer,Take care of plagiarism,Answer completely,You will get up vote for sure.
arrow_forward
Question 2
In a test measuring the life span of a certain brand of tire, 100 tires are tested. The results
showed an average lifetime of 50,000 miles, with a standard deviation of 5,000 miles. One
hundred thousand of these tires have been sold and are on the road.
(a) After what mileage would you expect 10% of the tires to have worn out?
(b) How many tires are expected to wear out between 60,000 and 70,000 miles?
(c) How many tires are expected to have a life of less than 20,000 miles?
(d) What are your major assumptions in the calculations for (a), (b) and (c)?
(e) Estimate the 95% confidence interval on the population mean.
arrow_forward
Match the terms precision and accuracy to their definitions.
Precision
2
Accuracy
Match each of the options above to the Items below.
The number of significant digits or the closeness of different measurements of the same quantity to each other.
The closeness of the measured or calculated value to the true value
1
2
arrow_forward
University of Babylon
Collage of Engineering\Al-Musayab
Department of Automobile
Engineering
Under Grad/Third stage
Notes:
1-Attempt Four Questions.
2- Q4 Must be Answered
3-Assume any missing data.
4 تسلم الأسئلة بعد الامتحان مع الدفتر
Subject: Mechanical
Element Design I
Date: 2022\01\25
2022-2023
Time: Three Hours
Course 1
Attempt 1
Q1/ Design a thin cylindrical pressure tank (pressure vessel) with hemispherical ends to the
automotive industry, shown in figure I below. Design for an infinite life by finding the
appropriate thickness of the vessel to carry a sinusoidal pressure varied from {(-0.1) to (6) Mpa}.
The vessel is made from Stainless Steel Alloy-Type 316 sheet annealed. The operating
temperature is 80 C° and the dimeter of the cylinder is 36 cm. use a safety factor of 1.8.
Fig. 1
(15 Marks)
Q2/ Answer the following:
1- Derive the design equation for the direct evaluation of the diameter of a shaft to a desired
fatigue safety factor, if the shaft subjected to both fluctuated…
arrow_forward
The data shown in Table 1 are the deviations from nominal diameter for holes drilled in a
carbon-fiber composite material used in aerospace manufacturing. The values reported are
deviations from nominal in ten-thousandths of an inch.
(a) Set up and R charts on the process. Is the process in statistical control?
(b) Estimate the process standard deviation using the range method.
(c) If specifications are at nominal ±100, what can you say about the capability of this
process? Calculate the PCR Cp
Table 1. Hole Diameter Data
Sample
Number X1 X2
1
2
3
4
5
6
7
8
9
10
11
12
14
15
16
17
18
19
20
-30 +50
0
-50
-10 -10
+20
-40
0
0
-30
0
+20
0
0
+70
0
+10
+40
+30
+30
+30
+10
0
+20
+50
+50
-20
+50 -60
+10
0
0
+20
+10 -20
-10
0
+20
X3 X4 X5
-30
+20
+30
+50
+40
+20
+30
+20
+30
+20
+30
0
+20
+10 +40
+10 +10
-10 +50
-10 -30
-10
+50
+40
0
+30
-10
0
+30
+30
-20
+50 +30 +10
+40
+20
0 +30
+10 +30
-20 +30
+30
+20
-20 +50
+20
+10
-40 +20
-20
-10
-10
0
-20
+10
+10 +50
0
+10
arrow_forward
Need help filling out the table with the data I have (1st row is fine but show work please)
arrow_forward
7
arrow_forward
Problem 3:
Insulation
To=1
Toowwww
Steam
Tx2
T₂ T3
www www
R₁ R₁ R₂
www.T
R₂
Steam at T1 = 320 °C flows in a cast iron pipe (k= 80 W/m. °C) whose inner and outer diameters are
5 cm = 0.05 m and D₂ = 5.5 cm = 0.055 m, respectively. The pipe is covered with 3-cm-thick glass wool
insulation with k = 0.05 W/m. °C. Heat is lost to surroundings at T2 = 5 °C by natural convection and
radiation, with a combined heat transfer coefficient of h₂ = 18 W/m². °C. Taking the heat transfer coefficient
inside the pipe to be h₁ = 60 W/m². °C, determine the temperature drops across the pipe and the insulation.
The determination is based on a unit length of the pipe (L = 1 m).
Assumptions
1. Heat transfer is one-dimensional since there is no indication of any change with time.
2.
Heat transfer is one-dimensional since there is thermal symmetry about the centreline and no
variation in the axial direction.
3. Thermal conductivities are constant.
4. The thermal contact resistant at the interface is…
arrow_forward
You are a biomedical engineer working for a small orthopaedic firm that fabricates rectangular shaped fracture
fixation plates from titanium alloy (model = "Ti Fix-It") materials. A recent clinical report documents some problems with the plates
implanted into fractured limbs. Specifically, some plates have become permanently bent while patients are in rehab and doing partial
weight bearing activities.
Your boss asks you to review the technical report that was generated by the previous test engineer (whose job you now have!) and used to
verify the design. The brief report states the following... "Ti Fix-It plates were manufactured from Ti-6Al-4V (grade 5) and machined into
solid 150 mm long beams with a 4 mm thick and 15 mm wide cross section. Each Ti Fix-It plate was loaded in equilibrium in a 4-point bending
test (set-up configuration is provided in drawing below), with an applied load of 1000N. The maximum stress in this set-up was less than the
yield stress for the Ti-6Al-4V…
arrow_forward
SEE MORE QUESTIONS
Recommended textbooks for you
Elements Of Electromagnetics
Mechanical Engineering
ISBN:9780190698614
Author:Sadiku, Matthew N. O.
Publisher:Oxford University Press
Mechanics of Materials (10th Edition)
Mechanical Engineering
ISBN:9780134319650
Author:Russell C. Hibbeler
Publisher:PEARSON
Thermodynamics: An Engineering Approach
Mechanical Engineering
ISBN:9781259822674
Author:Yunus A. Cengel Dr., Michael A. Boles
Publisher:McGraw-Hill Education
Control Systems Engineering
Mechanical Engineering
ISBN:9781118170519
Author:Norman S. Nise
Publisher:WILEY
Mechanics of Materials (MindTap Course List)
Mechanical Engineering
ISBN:9781337093347
Author:Barry J. Goodno, James M. Gere
Publisher:Cengage Learning
Engineering Mechanics: Statics
Mechanical Engineering
ISBN:9781118807330
Author:James L. Meriam, L. G. Kraige, J. N. Bolton
Publisher:WILEY
Related Questions
- Truncation errors are increased as the round-off errors are decreased.Group of answer choices True False Say, you have a thermometer and you are checking the temperature of a body that has a temperature of 36o Using your thermometer five times, it gives you the following measurements: 29oC, 29.2oC, 29.3oC, 28.9oC, and 29.1oC. What can we conclude about the accuracy and the precision of the thermometer?Group of answer choices The thermometer is not accurate and not precise The thermometer is faulty. The thermometer is accurate and precise The thermometer is not accurate but precise. Say, you have a thermometer and you are checking the temperature of a body that has a temperature of 36o Using your thermometer five times, it gives you the following measurements: 36oC, 35.6oC, 36oC, 37oC, and 36.2oC. What can we conclude about the accuracy and the precision of the thermometer?Group of answer choices The thermometer is accurate and precise. The thermometer is accurate but not precise. The…arrow_forwardA new electronic component for aircraft is tested to withstand temperatures as low as -50°F. One engineer suggests that the device cannot actually stand temperatures that low. What are the Type I and Type II errors? Type I: The device CAN NOT withstand temperatures as low as -50F but the engineer incorrectly finds that it CAN.Type II: The device CAN withstand temperatures as low as -50F but the engineer incorrectly finds that it CAN NOT. Type I: The device CAN withstand temperatures as low as -50F but the engineer incorrectly finds that it CAN NOT.Type II: The device CAN NOT withstand temperatures as low as -50F but the engineer incorrectly finds that it CAN. Is the answer the first or second bullet point?arrow_forwardCan someone please help me to answer all of the following questions thank you!!arrow_forward
- Question 4arrow_forwardA. A student establishes the time constant of a temperature sensor by first holding it immersed in hot water and then suddenly removing it and holding it immersed in cold water. Several other students perform the same test with similar sensors. Overall, their results are inconsistent, with estimated time constants differing by as much as a factor of 1.2. Offer suggestions about why this might happen. Hint: Try this yourself and think about control of test conditions. B. Which would you expect to be better suited to measure a time-dependent temperature, a thermal sensor (e.g., a thermocouple) having a small diameter spherical bead or one having a large diameter spherical bead? Why?arrow_forwardI need help solving this problem.arrow_forward
- Chapter 12 - Lecture Notes.pptx: (MAE 272-01) (SP25) DY... Scoresarrow_forwardQuestion 2 You are a biomedical engineer working for a small orthopaedic firm that fabricates rectangular shaped fracture fixation plates from titanium alloy (model = "Ti Fix-It") materials. A recent clinical report documents some problems with the plates implanted into fractured limbs. Specifically, some plates have become permanently bent while patients are in rehab and doing partial weight bearing activities. Your boss asks you to review the technical report that was generated by the previous test engineer (whose job you now have!) and used to verify the design. The brief report states the following... "Ti Fix-It plates were manufactured from Ti-6Al-4V (grade 5) and machined into solid 150 mm long beams with a 4 mm thick and 15 mm wide cross section. Each Ti Fix-It plate was loaded in equilibrium in a 4-point bending test (set-up configuration is provided in drawing below), with an applied load of 1000N. The maximum stress in this set-up was less than the yield stress for the…arrow_forwardb) A linear variable displacement transducer (LVDT) senses displacement and indicates a voltage output, which is linear to the input. Figure 1 shows an LVDT setup used for static calibration. It uses a micrometre to apply the known displacement and a voltmeter for the output. i) Suggest the independent, dependent and extraneous variables in this calibration. ii) Indicate the basic functional elements of the device in Figure 1 in the form of block diagram. 10 Micrometer Fixed mounts Connector Excitation voltage in Figure 1 Primary coil xx_w_ LVDT E Output voltage 10.135 Voltmeter Secondary coil Corearrow_forward
- Do not provide handwritten solution,Maintain accuracy and quality in your answer,Take care of plagiarism,Answer completely,You will get up vote for sure.arrow_forwardQuestion 2 In a test measuring the life span of a certain brand of tire, 100 tires are tested. The results showed an average lifetime of 50,000 miles, with a standard deviation of 5,000 miles. One hundred thousand of these tires have been sold and are on the road. (a) After what mileage would you expect 10% of the tires to have worn out? (b) How many tires are expected to wear out between 60,000 and 70,000 miles? (c) How many tires are expected to have a life of less than 20,000 miles? (d) What are your major assumptions in the calculations for (a), (b) and (c)? (e) Estimate the 95% confidence interval on the population mean.arrow_forwardMatch the terms precision and accuracy to their definitions. Precision 2 Accuracy Match each of the options above to the Items below. The number of significant digits or the closeness of different measurements of the same quantity to each other. The closeness of the measured or calculated value to the true value 1 2arrow_forward
arrow_back_ios
SEE MORE QUESTIONS
arrow_forward_ios
Recommended textbooks for you
- Elements Of ElectromagneticsMechanical EngineeringISBN:9780190698614Author:Sadiku, Matthew N. O.Publisher:Oxford University PressMechanics of Materials (10th Edition)Mechanical EngineeringISBN:9780134319650Author:Russell C. HibbelerPublisher:PEARSONThermodynamics: An Engineering ApproachMechanical EngineeringISBN:9781259822674Author:Yunus A. Cengel Dr., Michael A. BolesPublisher:McGraw-Hill Education
- Control Systems EngineeringMechanical EngineeringISBN:9781118170519Author:Norman S. NisePublisher:WILEYMechanics of Materials (MindTap Course List)Mechanical EngineeringISBN:9781337093347Author:Barry J. Goodno, James M. GerePublisher:Cengage LearningEngineering Mechanics: StaticsMechanical EngineeringISBN:9781118807330Author:James L. Meriam, L. G. Kraige, J. N. BoltonPublisher:WILEY
Elements Of Electromagnetics
Mechanical Engineering
ISBN:9780190698614
Author:Sadiku, Matthew N. O.
Publisher:Oxford University Press
Mechanics of Materials (10th Edition)
Mechanical Engineering
ISBN:9780134319650
Author:Russell C. Hibbeler
Publisher:PEARSON
Thermodynamics: An Engineering Approach
Mechanical Engineering
ISBN:9781259822674
Author:Yunus A. Cengel Dr., Michael A. Boles
Publisher:McGraw-Hill Education
Control Systems Engineering
Mechanical Engineering
ISBN:9781118170519
Author:Norman S. Nise
Publisher:WILEY
Mechanics of Materials (MindTap Course List)
Mechanical Engineering
ISBN:9781337093347
Author:Barry J. Goodno, James M. Gere
Publisher:Cengage Learning
Engineering Mechanics: Statics
Mechanical Engineering
ISBN:9781118807330
Author:James L. Meriam, L. G. Kraige, J. N. Bolton
Publisher:WILEY