REDOofSpectroscopy II Lab Report
pdf
keyboard_arrow_up
School
University of Akron *
*We aren’t endorsed by this school
Course
154
Subject
Chemistry
Date
Feb 20, 2024
Type
Pages
8
Uploaded by MateNeutronSkunk24
Moran-1
Spectroscopy II Lab Report
Chemistry 154.003
Exp. 3: Spectroscopy II
Kaylee Moran
February 8, 2024
Partner: Sydney Stacklin
TA: Sravya
Objective
The goal of this experiment is to figure out the concentrations of two species (Nickel, Ni, and Cobalt, Co) present in an unknown solution #282 and both of the species are colored.
Experimental Procedure
The procedure used was based on that in Semi-Quantitative Experiments for General Chemistry, The University of Akron, Spring 2020 Edition (course lab manual)
. Two 25mL burets and two cuvettes were checked out of the stockroom. Both cuvettes were matched using deionized water roughly 2-3 millimeters above the minimum fill mark, keeping the second sample cuvette within 1% above or below the first reference cuvette. 40 mL of both nickel and cobalt and were obtained and the maximum wavelength (λmax) and molarity (M) of both were recorded in the lab notebook. Dilutions were prepared for both nickel and cobalt according to Table 1
and were put into clean, dry, and labeled test tubes and then stirred with a clean, dry stirring rod. Each test tube was poured in turn into the sample cuvette and placed into spectrophotometer at nickel’s λmax and cobalt’s λmax. %T was recorded for each test tube dilution and then converted to absorbance (A). A Beer-Lambert plot (
Figure 1 and
Figure 2
) was crafted for nickel and cobalt (A vs concentration) to determine their respective molar absorptivity (
ε
Ni
and ε
Co
).
Moran-2
The spectrophotometer was changed to nickel’s λmax (390nm) which is λ
1
. Cuvette with pure, undiluted cobalt was placed into spectrophotometer and %T was recorded. The spectrophotometer was changed to cobalt’s λmax (510nm), which is λ
2
. A cuvette with pure, undiluted nickel was placed into spectrophotometer and %T was recorded. The unknown sample #282 was then placed into the sample cuvette and tested in spectrophotometer at 390 nm, resulting in a value of %T. The %T needed to be between 10-90%, so the unknown sample #282 was diluted, which is shown in Table 4
. 20 mL water and 2 mL unknown solution #282 was diluted, stirred, and %T was taken at 390 nm, this time resulting in a %T value of 56.1%T, which was in the accepted range. The diluted #282 sample was taken at 510 nm, resulting in a %T value of 66.0%. Both of these recorded %T values were then converted to A and then used to find the concentration of nickel and cobalt in the undiluted unknown sample #282.
Sample Calculations
Converting from %T to A:
A= 2 – log (%T)
For 15mL nickel at 390nm, %T= 32.7%
A= 2 – log (32.7)
A= 0.4855
Determining the Concentration of a Diluted Standard Sample
For 12mL Standard solution sample:
M
1
V
1
= M
2
V
2
M
2
= concentration (C)
0.1M
×
12mL
= M
2
×
15mL
M
2
=
0.1M×12mL
15mL
M
2
= 0.08
Moran-3
One-Point Determinations
To find ε
Ni2
:
A =
𝛆
𝐥𝐂
0.0048= ε
× 1𝑐𝑚 × 0.1
𝛆
=
0.0048
1×0.1
ε
Ni2=0.048
Determining Concentration in Unknown Sample
For Cobalt in Unknown:
C
Co
= A
1
/
ε
Ni1
ℓ
– A
2
/
ε
Ni2
ℓ
÷
ε
Co1
ℓ
/
ε
Ni1
ℓ - ε
Co2
ℓ/
ε
Ni2
ℓ
C
Co
= 0.2708/4.7915 – 0.1801/0.048 ÷
0.250/4.7915 – 5.026/0.048
C
Co
= -3.695/-104.656
C
Co
= 0.035311 M
For Nickel in Unknown:
A
1
=
ε
Ni1
ℓ
C
Ni + ε
Co1
ℓ
C
C0
0.2708 = 4.7915
×
1
×
C
Ni + 0.250
×
1
×
0.035311
0.2708 =4.7915
×
C
Ni + 0.00882775
0.26197= 4.7915
×
C
Ni
To find ε
Co1:
A =
𝛆
𝐥𝐂
0.0250=
ε × 1𝑐𝑚 × 0.1
ε
=
0.0250
1×0.1
ε
Co
1=0.250
Your preview ends here
Eager to read complete document? Join bartleby learn and gain access to the full version
- Access to all documents
- Unlimited textbook solutions
- 24/7 expert homework help
Moran-4
C
Ni
= 0.05467 M
Determining Concentration of Unknown in Undiluted Sample
?
1
𝑉
1
=
?
2
𝑉
2
?
1
× 2𝑚?
= 0
.035311?
× 22𝑚?
?
1
=
0
.035311? ×22𝑚? 2𝑚?
?
1
=
0.388M
Results
From the bottle of Nickel Sample:
[Ni] = 0.1 M
λ
1
=390 nm
From the bottle of Cobalt Sample:
[Co] = 0.1 M
λ
2
=510 nm
Table 1
Dilutions for each labeled test tube. These exact dilutions were done for both nickel and cobalt.
Moran-5
Standard Standard (mL) Water (mL) %T A C 390 nm 15.00 0.00 32.7 0.4855 0.1 12.00 3.00 45.2 0.3448 0.08 9.00 6.00 60.8 0.2161 0.06 6.00 9.00 67.4 0.1713 0.04 3.00 12.00 80.7 0.0931 0.02 510 nm 15.00 0.00 98.9 0.0048 0.1 Table 2
%T, A, and C values for Nickel at 390 nm (λ
1) and 510 nm (
λ
2)
.
Figure 1.
Beer
‐
Lambert plot of nickel at its λmax value of 390 nm. The molar absorptivity
at 390 nm was determined from the best fit line.
From the best fit line of the Beer Lambert plot, the molar absorptivity (
ε
Ni
)
of nickel at 390 nm was determined to be 4.7915
𝑴
−𝟏
𝒄𝒎
−𝟏
.
Standard (mL) Water (mL) %T A C 510 nm 15.00 0.00 32.7 0.4855 0.1 12.00 3.00 50.9 0.2933 0.08 9.00 6.00 61.2 0.2132 0.06 6.00 9.00 78.8 0.1035 0.04 3.00 12.00 83.6 0.0778 0.02 390 nm 15.00 0.00 94.4 0.0250 0.1 Table 3
%T, A, and C values for Cobalt at 390 nm (λ
1) and 510 nm (
λ
2)
.
y = 4.7915x - 0.0253
R² = 0.9599
0
0.1
0.2
0.3
0.4
0.5
0.6
0
0.02
0.04
0.06
0.08
0.1
0.12
Absorbance
Concentration (M)
Beer-Lambert Plot of Nickel at 390 nm
Moran-6
Figure 2.
Beer
‐
Lambert plot of cobalt at its λmax value of 510 nm. The molar absorptivity
at 510 nm was determined from the best fit line.
From the best fit line of the Beer Lambert plot, the molar absorptivity (
ε
Co
)
of cobalt at 510 nm was determined to be 5.026
𝑴
−𝟏
𝒄𝒎
−𝟏
.
y = 5.026x - 0.0669
R² = 0.9302
0
0.1
0.2
0.3
0.4
0.5
0.6
0
0.02
0.04
0.06
0.08
0.1
0.12
Absorbance
Concentration (M)
Beer-Lambert Plot of Cobalt at 510 nm
Your preview ends here
Eager to read complete document? Join bartleby learn and gain access to the full version
- Access to all documents
- Unlimited textbook solutions
- 24/7 expert homework help
Moran-7
Unknown Sample
Unknown #282
UNK #282 Dilution Unknown (mL) V
initial
0.00 V
final
2.00 V
total
2.00 Water (mL) V
initial
0.00 V
final
20.00 V
total
20.00 V
diluted unknown
(mL) 22.00 %T at 390nm 0.1 53.6 A at 390 nm 3.00 0.2708 %T at 510 nm 66.0 A at 510 nm 0.1805 The concentration of cobalt in unknown #282 was determined to be 0.388M and the concentration of nickel in the unknown #282 was determined to be 0.6014M.
Discussion
In this experiment, the values of C
Co
and C
Ni
were determined for Sample #282. Figure 1
presented the molar absorptivity of Nickel at 390nm, determined to be 4.7915
?
−1
𝑐𝑚
−1
, which is ε
Ni1. Figure 2
presented the molar absorptivity of Cobalt at 510nm, determined to be 5.026
?
−1
𝑐𝑚
−1
, which is ε
Co2. ε
Co1 and ε
Ni2 were found using one-point determinations found on page three. ε
Co1 was determined to be 0.250
?
−1
𝑐𝑚
−1
and ε
Ni2 was determined to be 0.048
?
−1
𝑐𝑚
−1
. Absorptivity of unknown #282 undiluted sample at 390nm was found to be Table 4
The initial, final, and total volumes of water and UNK #282 for the dilution, as well as the %T and A value for each solution. The dilution yielded a usable %T value.
Moran-8
0.2708, which is A
1
. Absorptivity of unknown #282 undiluted sample at 510nm was found to be 0.1801, which is A
2
. Sample
#282’s C
Co
was found to be 0.0353M by using ε
Co1, ε
Co2, ε
Ni1,
ε
Ni2, A
1, and A
2 values previously stated.
The resulting value of 0.0353M was then used to find the concentration of nickel (C
Ni
) in the unknown #282 sample, resulting in a value of 0.05467M. C
Co and C
Ni values were then used to determine the concentration of Cobalt and Nickel in the unknown undiluted sample by using the formula of ?
1
𝑉
1
=
?
2
𝑉
2
. M
1
of Cobalt was solved for by using 2mL for V
1
, M
2
was C
co
value of 0.0353, and 22mL for V
2
, resulting in a concentration of 0.388M. M
1
of Nickel was solved for by using 2mL for V
1
, M
2
was C
Ni
value of 0.05467, and 22mL for V
2
, resulting in a concentration of 0.6014M. Therefore, the concentration of cobalt in the unknown #282 undiluted sample was determined to be 0.388M and the concentration of nickel was determined to be 0.6014M.
Several sources of error could have influenced this experiment to varying extents. Achieving complete drying of the cuvettes after water rinsing posed a significant challenge. The residual water in the cuvettes could potentially dilute any subsequently added samples, resulting in elevated %T values and reduced A values. If such dilution occurred during the unknown sample's preparation, the experimental A value and concentration might register as lower than the actual values. Despite efforts to rinse cuvettes with the measured solution, this phenomenon could still exert an impact, ultimately leading to incorrect concentration values in the unknown. Additionally, the presence of scratches, smudges, or internal dust within the cuvette was a plausible source of error. In these scenarios, the introduction of extra solids or imperfections in the glass could scatter the initial radiation before reaching the solution in the cuvette. Consequently, the amount of transmitted radiation might be diminished, leading to higher A values and consequently, higher concentrations.
Related Documents
Related Questions
Chemistry
Instrumental Analysis Class
The experiment is OES: Cleaning Validation of removal of pharmaceutical drugs and calibration standards by fluoresence spectroscopy
- Why the instrument use two detectors and monochromators? What type of monochromators are used in our instrument?
- Do you think that a similar experiment can be prepared for the analysis of naproxen and benzoic acid? Explain.
arrow_forward
B. Quantitative Spectroscopy
Prepare two graphs, %T vs. Concentration and Absorbance vs. Concentration, using the data for the cobalt standards. The concentrations of Cobalt standards 2-4 can be calculated using simple dilution calculations. Concentration should be on the x-axis for each of these graphs. The first should clearly show some curvature while the second should be a linear curve.
Using the absorbance graph you can determine the concentration of the unknown. Take the absorbance value for the unknown and find that value on the y-axis. Draw a horizontal line from that point on the y-axis to the plotted line. From the point where the horizontal line intersects the plotted line, draw a vertical line straight down to the x-axis. Record the value from the point where you hit the x-axis as the unknown concentration.
arrow_forward
Do
arrow_forward
PRE-LAB QUESTIONS
Experiment 121
Complete these questions prior to attending lab. Some of the results will be useful in conducting
the experiment, so you should record those results in the appropriate section of the lab as well.
1. In Part A, step 1, you are directed to set up the
spectrophotometer for measurement of the
FeSCN2+ product ion.
a. Based on the spectra given here, what
would be an appropriate wavelength for the
measurement? Review Appendix I if
necessary to again familiarize yourself with
the issues related to this decision.
b. Briefly explain your choice.
0.5
A 0.4
0.3
Absorbance A
Absorbance spectra for Fe³+ and FeSCN²+
0.2
0.1
0
400
0.02 M Fe³+ (dissolved in HNO3)
E121-11
0.000048 M FeSCN²+
500
600
wavelength (nm)
700
2. In Part A, step 4, you will experimentally determine the value of the molar absorptivity, &, for
FeSCN2+ at your selected wavelength. Using the Beer-Lambert law, calculate an approximate
value for & based on the spectrum of 0.000048 M FeSCN2+ provided…
arrow_forward
Which of the following explains why a more concentrated dye solution has a higher absorbance?
A. There are more dye molecules in the solution so more isotopes reach the detector.
b. There are more isotopes in solution to emit photons.
c. There are more dye molecules in solution to emit isotopes.
d. There are more isotopes in solution to emit protons.
e. There are more dye molecules in the solution to absorb photons.
arrow_forward
A calibration curve is made for the absorbance as a function of concentration of blue dye (in mg/L). The slope of the curve is 0.131; the y-intercept is -0.01869; the correlation is 0.993. If a consumer sample has an absorbance of 0.120 on the same instrument, what is its concentration of blue dye, in mg/L?
a. 0.034
b. 0.935
c. 0.993
d. 1.06
arrow_forward
Determine the standard concentration of A through E and then the absorbance. I just need clarification on if I add together the two ml values given.
arrow_forward
I need help calculating the molar absorption please!
arrow_forward
What is the purpose of cobalt blue glass in spectroscopy?
arrow_forward
An exploration of any one application of Spectroscopy (Absorption or Emission), Mass Spectrometry (MS) - (ANY kind of mass spec)
Pick one location/project/bit of research where the data is used and explore. Discuss on what is measured (what instrument) and why and how. You don't have to fully teach the instrument but spend a paragraph giving the very basics of what is measured and how. And discuss the application you discovered - just find a place the measurement is used and explain what they do - what they measure and when it tells them.
arrow_forward
Which of the following is not a step in Atomic Absorption Spectroscopy (AAS)?
A.
Atoms absorb light.
B.
A solution is vaporized
C.
A calibration curve is constructed.
D.
Particles are adsorbed onto a stationary phase.
arrow_forward
A colored ion solution has a concentration of 0.200 M with a measured absorbance A = 0.880. Another ion solution made of the same chemicals has an absorbance A = 0.172. What is the concentration of this unknown sample solution?
arrow_forward
.Q1. To determine the concentrations (mol/L) of Co(NO3)2 (A) and Cr(NO3)3(B) in an unknown sample, the following representative absorbance data wereobtained.A (mol/L) B (mol/L) 510nm 575nm5×10−1 0 0.714 0.0970 6×10−2 0.298 0.757Unknown Unknown 0.671 0.330Measurements were made in 1.0 cm glass cells.i. Calculate the four molar absorptivities: ∈A(510), ∈A(575), ∈B(510) and ∈B(575).ii. Calculate the molarities of the two salts A and B in the unknown.
arrow_forward
Explain tropylium ion formation in Electron Ionization Mass Spectroscopy (EI-MS). Please answer at your own easy words. Answer should be to the point.
arrow_forward
26. What exactly is flame atomic absorbtion spectrometer?
arrow_forward
Solutions that contain the FeSCN2+ ion are blood red (you will see this at the end of the second quarter of lab). Consider the regions of the visible spectrum in which the compound should absorb light and prepare a rough sketch of absorbance vs. wavelength for each solution. Label λ max on the graph. (Is this question asking for the graph of my data at the end)
arrow_forward
Which type of mass spectrometer ionization results in the greatest amount of fragmentation
A.atmospheric pressure chemical ionization
B.electrospray
C.chemical ionization
D.electron impact
arrow_forward
Which of the following statements are true of atomic emission spectroscopy (AES)?
I. Atoms are promoted to an excited state with a laser.
II. Atoms are promoted to an excited state by gaining energy from collisions with other atoms or from the high thermal energy of the flame.
III. The intensity of emitted light is proportional to analyte concentration.
IV. Plasma often replaces flame in AES
A.
III and IV
B.
I, III, and IV
C.
II, III, and IV
D.
II and III
arrow_forward
How does spectroscopy help to identify an element?
arrow_forward
Application Questions: Answer the following questions. Show all work for calculations and
include units and the correct number of significant figures.
1. A solution of cobalt (III) chloride was analyzed using a UV/Vis spectrometer and produced the
spectra shown in Fig. 4. What is the best wavelength to use for analyzing the samples?
Best wavelength to analyze sample:
Absorbance
0.600
0.500
0.400
0.300
0.200
0.100
0.000
375
400
450
425
475
500 525 550
Wavelength (nm)
Molarity of
FD&C #2
0.000
0.115
0.225
0.350
575
Figure 4. UV/Vis spectra of a cobalt (III) chloride solution for determination of the best
wavelength for analysis.
8
600 625
2. Create a graph in Excel and add a linear trendline using the data given below. Calculate the
molarity of a solution (to 3 significant digits) with an absorbance of 0.576 using the trendline
equation from this graph. Show your work in the box below. DO NOT ATTACH THE
FIGURE.
Abs
0.002
0.187
0.442
0.699
650
675
arrow_forward
Please answer all and post below the answer the source from where you got the answer
thank you!
arrow_forward
SEE MORE QUESTIONS
Recommended textbooks for you

Macroscale and Microscale Organic Experiments
Chemistry
ISBN:9781305577190
Author:Kenneth L. Williamson, Katherine M. Masters
Publisher:Brooks Cole
Related Questions
- Chemistry Instrumental Analysis Class The experiment is OES: Cleaning Validation of removal of pharmaceutical drugs and calibration standards by fluoresence spectroscopy - Why the instrument use two detectors and monochromators? What type of monochromators are used in our instrument? - Do you think that a similar experiment can be prepared for the analysis of naproxen and benzoic acid? Explain.arrow_forwardB. Quantitative Spectroscopy Prepare two graphs, %T vs. Concentration and Absorbance vs. Concentration, using the data for the cobalt standards. The concentrations of Cobalt standards 2-4 can be calculated using simple dilution calculations. Concentration should be on the x-axis for each of these graphs. The first should clearly show some curvature while the second should be a linear curve. Using the absorbance graph you can determine the concentration of the unknown. Take the absorbance value for the unknown and find that value on the y-axis. Draw a horizontal line from that point on the y-axis to the plotted line. From the point where the horizontal line intersects the plotted line, draw a vertical line straight down to the x-axis. Record the value from the point where you hit the x-axis as the unknown concentration.arrow_forwardDoarrow_forward
- PRE-LAB QUESTIONS Experiment 121 Complete these questions prior to attending lab. Some of the results will be useful in conducting the experiment, so you should record those results in the appropriate section of the lab as well. 1. In Part A, step 1, you are directed to set up the spectrophotometer for measurement of the FeSCN2+ product ion. a. Based on the spectra given here, what would be an appropriate wavelength for the measurement? Review Appendix I if necessary to again familiarize yourself with the issues related to this decision. b. Briefly explain your choice. 0.5 A 0.4 0.3 Absorbance A Absorbance spectra for Fe³+ and FeSCN²+ 0.2 0.1 0 400 0.02 M Fe³+ (dissolved in HNO3) E121-11 0.000048 M FeSCN²+ 500 600 wavelength (nm) 700 2. In Part A, step 4, you will experimentally determine the value of the molar absorptivity, &, for FeSCN2+ at your selected wavelength. Using the Beer-Lambert law, calculate an approximate value for & based on the spectrum of 0.000048 M FeSCN2+ provided…arrow_forwardWhich of the following explains why a more concentrated dye solution has a higher absorbance? A. There are more dye molecules in the solution so more isotopes reach the detector. b. There are more isotopes in solution to emit photons. c. There are more dye molecules in solution to emit isotopes. d. There are more isotopes in solution to emit protons. e. There are more dye molecules in the solution to absorb photons.arrow_forwardA calibration curve is made for the absorbance as a function of concentration of blue dye (in mg/L). The slope of the curve is 0.131; the y-intercept is -0.01869; the correlation is 0.993. If a consumer sample has an absorbance of 0.120 on the same instrument, what is its concentration of blue dye, in mg/L? a. 0.034 b. 0.935 c. 0.993 d. 1.06arrow_forward
- An exploration of any one application of Spectroscopy (Absorption or Emission), Mass Spectrometry (MS) - (ANY kind of mass spec) Pick one location/project/bit of research where the data is used and explore. Discuss on what is measured (what instrument) and why and how. You don't have to fully teach the instrument but spend a paragraph giving the very basics of what is measured and how. And discuss the application you discovered - just find a place the measurement is used and explain what they do - what they measure and when it tells them.arrow_forwardWhich of the following is not a step in Atomic Absorption Spectroscopy (AAS)? A. Atoms absorb light. B. A solution is vaporized C. A calibration curve is constructed. D. Particles are adsorbed onto a stationary phase.arrow_forwardA colored ion solution has a concentration of 0.200 M with a measured absorbance A = 0.880. Another ion solution made of the same chemicals has an absorbance A = 0.172. What is the concentration of this unknown sample solution?arrow_forward
arrow_back_ios
SEE MORE QUESTIONS
arrow_forward_ios
Recommended textbooks for you
- Macroscale and Microscale Organic ExperimentsChemistryISBN:9781305577190Author:Kenneth L. Williamson, Katherine M. MastersPublisher:Brooks Cole

Macroscale and Microscale Organic Experiments
Chemistry
ISBN:9781305577190
Author:Kenneth L. Williamson, Katherine M. Masters
Publisher:Brooks Cole