Interpretation:
The type of mass spectrum used for molecular mass of compound and reason for using this mass spectrum is to be stated.
Concept introduction:
The mass spectrum of a compound is obtained by the ionization of the compound. The ionization can be done by using electron ionization or chemical ionization. The type in which the spectrum of compound obtained is simple and the process of the method is easy.

Answer to Problem 20.19QAP
The chemical ionization of the mass spectrum is used for molecular mass of a compound.
Explanation of Solution
The mass spectrum obtained by chemical ionization is used for molecular mass of compound because the molecular structure of the compound is disturbed in electron ionization. In electron ionization bombardment of electrons and photons is used to produce secondary ions. The energy consumed in electron ionization is high and the exact molecular structure is not obtained. If the molecular structure of the compound is different, then the mass spectrum obtained is also different and the molecular mass of the compound cannot be determined.
The chemical ionization process to obtain a spectrum of a compound is simple and consumes low energy. Thus, the chemical ionization of the mass spectrum is used for molecular mass of a compound.
(b)
Interpretation:
The type of mass spectrum used for the chemical structure of the compound and reason for using this is to be stated.
Concept introduction:
The chemical structure of the compound consists of the different orientation of atoms in a compound. The analysis of every atom is done to obtain a chemical structure of compound because a particular atom behaves differently with a different atom in the compound. The method used for the mass spectrum for the chemical structure of the compound should be such that the orientation of every atom can be analyzed separately.

Answer to Problem 20.19QAP
The electron ionization of the mass spectrum is used for the chemical structure of the compound.
Explanation of Solution
The electron ionization is used in mass spectrum for chemical structure of compound because every atom is analysed by ionizing differently and spectrum of chemical structure is very simple.
(c)
Interpretation:
The flight time for ion in TOF mass analyzer is to be stated.
Concept introduction:
In TOF mass analyzer positive ions are produced by bombarding the electrons or photons on molecules of a compound. These produced ions then accelerated in the drift tube due to the electric field applied. The ions in the tube are separated on the basis of their masses. The kinetic energies of all the ions in the tube are the same because all ions are under the effect of the same electric field. The velocities of ions are in an inverse of their masses.

Answer to Problem 20.19QAP
The flight time for ion in TOF mass analyzer is
Explanation of Solution
Write the expression of flight time for ion.
Here, the flight time is
Substitute
(d)
Interpretation:
The derivation of an equation for the differences in flight times as a function of two masses, charges and accelerating voltage is to be stated.
Concept introduction:
The ion in the tube of mass analyzer consists of kinetic energy and potential energy. The potential energy is due to charged electrons and kinetic energy due to the mass of ions. The difference in flight times is obtained by finding the flight time of each ion separately.

Answer to Problem 20.19QAP
The equation for the differences in flight times as a function of two masses, charges and accelerating voltage is
Explanation of Solution
Write the expression for potential energy of charged electrons
Here, charge of electron is
Write the expression of kinetic energy of ions
Here, mass of ions is
The potential energy of electrons and kinetic energy of ions is same in TOF mass analyzer. From Equation (II) and Equation (III).
Write the expression for flight time
Here, length of tube is
Substitute
Write the expression of the flight time for ion of mass to charge ratio
Write the expression of the flight time for ion of mass to charge ratio
Write the expression for difference in flight time of ions
Substitute
(e)
Interpretation:
The difference between flight time for ions of mass to charge ratios
Concept introduction:
The difference in flight time for ions depends on mass of ions, charge of electrons and accelerating voltage in the tube. It is given by the formula

Answer to Problem 20.19QAP
The difference between flight time for ions of mass to charge ratios
Explanation of Solution
Write the expression of flight time of ions
Here, the mass of ions are
Substitute
Write the expression of flight time of ions
Substitute
Write the expression for differences in flight time
Substitute
(f)
Interpretation:
The type of ionization source in tendom mass spectrometer and reason for using this is to be stated.
Concept introduction:
The mass spectrum is the plot between the relative abundance and mass to charge ratio. The mass spectrum is different for the same substance. The relative abundance represents the y-axis and the mass to charge ratio represents the x-axis. The heights of the peaks represent the relative intensity. The peak with the highest value in the plot is called the base peak. All the calculations are made with respect to this base peak. The structural information can be obtained by fragmentation of molecules in the sample.

Answer to Problem 20.19QAP
The electronic ionization is used to get the structural information of the compound because the single unpaired electron is obtained by electronic ionization.
Explanation of Solution
The fragmentation pattern is obtained by passing the vaporized organic sample from the ionization chamber. This vaporized organic sample is then bombarded by the stream of electrons and positive ions are obtained. This positive ion is called molecular ion and consists of a single unpaired electron. This single unpaired electron is obtained by the electronic ionization process. The peak in the mass spectra with value
(g)
Interpretation:
The types of mass spectra obtained from the MS/MS experiment is to be stated.
Concept introduction:
The mass spectrometry is a technique used to find the structure and chemical properties of different molecules. In the process of mass spectrometry, different molecules are converted into ions. The ions are separated by the mass to charge ratio and detected by relative abundance in the sample. The mass spectrum of ion is obtained by plotting the mass to charge ratio and relative abundance of the ion in a sample.

Answer to Problem 20.19QAP
The type of mass spectra obtained from MS/MS experiment by:
(1) Holding the first analyzer constant and scanning the second analyzer is product ion spectra.
(2) Scanning both analyzers with a small
(3) Scanning the first analyzer and holding the second analyzer constant is precursor ion spectra.
(4) Scanning the second mass analyzer for every mass-selected by the first mass analyzer is three-dimensional spectra.
Explanation of Solution
(1) The ion spectra obtained in the MS/MS experiment is very different from each other for any compound. The product ion spectra of the compound are obtained in the MS/MS experiment by holding the first analyzer constant and scanning the second analyzer.
(2).When the difference between mass to charge ratio to different ions is very small, by scanning both analyzers, then ions are detected by analyzing the neutral loss spectrum.
(3) The compounds with the same products in ions are identified by determining the concentration of ions in each compound. If ions are detected by scanning the first mass analyzer and keeping the second mass analyzer constant, then precursor ion spectra are obtained in the MS/MS experiment.
(4) If ions are detected by selecting ions in the first mass analyzer and scanned by the second mass analyzer, then the three-dimensional spectrum is obtained in the MS/MS experiment.
Want to see more full solutions like this?
- What are the products of the following acetal hydrolysis? Please draw a skeletal line structure and include a detailed explanation and drawing of how the mechanism proceeds. Please include any relevant information that is needed to understand the process of acetal hydrolysis.arrow_forwardWhat would happen if you added the HCI to the Grignard reagent before adding benzophenone? Draw a reaction mechanism to support your answer.arrow_forwardAt 300 K, in the decomposition reaction of a reactant R into products, several measurements of the concentration of R over time have been made (see table). Calculate the order of the reaction. t/s [R]/ (mol L-1) 0 0,5 171 0,16 720 0,05 1400 0,027arrow_forward
- Write the correct IUPAC names of the molecules in the picturearrow_forwardHow many grams of solid NaCN have to be added to 1.5L of water to dissolve 0.18 mol of Fe(OH)3 in the form Fe(CN)63 - ? ( For simplicity, ignore the reaction of CN - ion with water) Ksp for Fe(OH)3 is 2.8E -39, and Kform for Fe(CN)63 - is 1.0E31arrow_forwardDraw the most stable chair conformation of 1-ethyl-1-methylcyclohexane, clearly showing the axial and equatorial substituents. [4] Draw structures corresponding to the following IUPAC name for each of the following compounds; [5] i) 4-Isopropyl-2,4,5-trimethylheptane ii) trans-1-tert-butyl-4-ethylcyclohexane iii) Cyclobutylcycloheptane iv) cis-1,4-di-isopropylcyclohexane (chair conformation) v) 3-Ethyl-5-isobutylnonanearrow_forward
- Draw and name molecules that meet the following descriptions; [4] a) An organic molecule containing 2 sp2 hybridised carbon and 1 sp-hybridised carbon atom. b) A cycloalkene, C7H12, with a tetrasubstituted double bond. Also answer question 2 from the imagearrow_forwardH 14. Draw the line angle form of the following molecule make sure you use the proper notation to indicate spatial positioning of atoms. F F H 15. Convert the following condensed form to line angle form: (CH3)3CCH2COCH2CON(CH2CH3)2arrow_forwardIn a reaction between two reactants A and B, the half-life is the same for both only if(A) the stoichiometry A:B is 1:1.(B) the stoichiometry A:B is 1:2 or 2:1.arrow_forward
- In a reaction between two reactants A and B, the half-life is the same for both.(1) Only if the stoichiometry A:B is 1:1.(2) If the initial quantities of A and B are in their stoichiometric ratios.arrow_forwardThere are 48 pairs of students in the following table. Each pair has quantitatively determined the mass of taurine in a 250 mL can of the popular energy drink marketed as “Munster” using High Performance Liquid Chromatography (HPLC). The class results are presented below: QUESTION: Calculate the measurement of uncertainty and provide the data in a spreadsheet table. Mass of Taurine (mg) Mass of Taurine (mg) (Table continued) 152.01 152.23 151.87 151.45 154.11 152.64 152.98 153.24 152.88 151.45 153.49 152.48 150.68 152.33 151.52 153.63 152.48 151.68 153.17 153.40 153.77 153.67 152.34 153.16 152.57 153.02 152.86 151.50 151.23 152.57 152.72 151.54 146.47 152.38 152.44 152.54 152.53 152.54 151.32 152.87 151.24 153.26 152.02 152.90 152.87 151.49 152.46 152.58arrow_forward1. Predict the organic product(s) of the following reactions. Assume excess of reagents unless otherwise noted. a) &l BH3 •THF b) 1) NaOH 2) H3O+ solve d) ala 1) EtMgBr 2) H3O+ e) H2N سكر CuLi NH2 1) SOCI2 2) EtMgBr 3) H3O+ NC H3O+ Δarrow_forward
- Principles of Instrumental AnalysisChemistryISBN:9781305577213Author:Douglas A. Skoog, F. James Holler, Stanley R. CrouchPublisher:Cengage LearningIntroduction to General, Organic and BiochemistryChemistryISBN:9781285869759Author:Frederick A. Bettelheim, William H. Brown, Mary K. Campbell, Shawn O. Farrell, Omar TorresPublisher:Cengage Learning