If a source contained 153 µCi of Bal133 (T, = 10.66 y) on June 1, 2005, what would be the 9. 10% acceptable activity range if the source is used for a dose calibrator accuracy test on June 1, 2007? (See Radiopharmacy Sections 1, 2, and 3.)
If a source contained 153 µCi of Bal133 (T, = 10.66 y) on June 1, 2005, what would be the 9. 10% acceptable activity range if the source is used for a dose calibrator accuracy test on June 1, 2007? (See Radiopharmacy Sections 1, 2, and 3.)