If a source contained 153 µCi of Bal133 (T, = 10.66 y) on June 1, 2005, what would be the 9. 10% acceptable activity range if the source is used for a dose calibrator accuracy test on June 1, 2007? (See Radiopharmacy Sections 1, 2, and 3.)

icon
Related questions
Question
If a source contained 153 uCi of Bal33 (T, = 10.66 y) on June 1, 2005, what would be the
9.
10% acceptable activity range if the source is used for a dose calibrator accuracy test on June
1, 2007? (See Radiopharmacy Sections 1, 2, and 3.)
Transcribed Image Text:If a source contained 153 uCi of Bal33 (T, = 10.66 y) on June 1, 2005, what would be the 9. 10% acceptable activity range if the source is used for a dose calibrator accuracy test on June 1, 2007? (See Radiopharmacy Sections 1, 2, and 3.)