Two units of DC voltmeters, A and B are used to measure a voltage from a resistor R2 as in Figure Q4 (b). These instrument specifications are stated as below: Meter A: Sensitivity S = 3 kN/V, Internal Resistance Rm = 0.5 kN, Range = 15 V Meter B: Sensitivity S = 15 kN/V, Internal Resistance Rm = 3 kN, Range = 20 V Ri=5kQ 30 V RF 25k2. Figure Q4(b) Calculate: (i) Voltage across R2 without any instrument connected. (ii) Voltage across R2 using instrument A. (iii) Voltage across R2 using instrument B. [2
Two units of DC voltmeters, A and B are used to measure a voltage from a resistor R2 as in Figure Q4 (b). These instrument specifications are stated as below: Meter A: Sensitivity S = 3 kN/V, Internal Resistance Rm = 0.5 kN, Range = 15 V Meter B: Sensitivity S = 15 kN/V, Internal Resistance Rm = 3 kN, Range = 20 V Ri=5kQ 30 V RF 25k2. Figure Q4(b) Calculate: (i) Voltage across R2 without any instrument connected. (ii) Voltage across R2 using instrument A. (iii) Voltage across R2 using instrument B. [2
Related questions
Question
![(b)
Two units of DC voltmeters, A and B are used to measure a voltage from a resistor R2 as in
Figure Q4 (b).
These instrument specifications are stated as below:
Meter A: Sensitivity S = 3 kN/V, Internal Resistance Rm = 0.5 kN, Range = 15 V
%3D
Meter B: Sensitivity S = 15 kN/V, Internal Resistance Rm = 3 kN, Range = 20 V
%3D
Ri=5kQ
30 V
RF 25k2
Figure Q4(b)
Calculate:
(i)
Voltage across R2 without any instrument connected.
(ii)
Voltage across R2 using instrument A.
(iii)
Voltage across R2 using instrument B.
(iv)
Absolute error for each measurement.](/v2/_next/image?url=https%3A%2F%2Fcontent.bartleby.com%2Fqna-images%2Fquestion%2Ff748742c-f959-46a5-8a1d-40867613353d%2F178c974e-49d7-4f83-b0cf-141145ff3bb1%2Ful30p5g_processed.jpeg&w=3840&q=75)
Transcribed Image Text:(b)
Two units of DC voltmeters, A and B are used to measure a voltage from a resistor R2 as in
Figure Q4 (b).
These instrument specifications are stated as below:
Meter A: Sensitivity S = 3 kN/V, Internal Resistance Rm = 0.5 kN, Range = 15 V
%3D
Meter B: Sensitivity S = 15 kN/V, Internal Resistance Rm = 3 kN, Range = 20 V
%3D
Ri=5kQ
30 V
RF 25k2
Figure Q4(b)
Calculate:
(i)
Voltage across R2 without any instrument connected.
(ii)
Voltage across R2 using instrument A.
(iii)
Voltage across R2 using instrument B.
(iv)
Absolute error for each measurement.