I U X2 More Formatting A A Q 4. A fluoride ion selective electrode was used to analyse for fluoride as part of the QC procedures during the production of Listerine Fluoride Defense Anticavity Mouthwash. The mouthwash contains 0.02% sodium fluoride which is equivalent to 90 ppm of fluoride. The QC specification for the mouthwash is 90 ± 5 ppm of fluoride. During production three standards and six periodic samples were analysed each hour. The mouthwash samples were diluted 100 fold before the pH and ionic strength were adjusted using an appropriate TISAB before analysis. The following data was obtained for the standards and the calibration curve is shown below. (i) (ii) Solution E (MV) 1.00 x 104 MF 40.5 5.00 x 10-5 MF 1.00 x 10-5 M F 60.2 100.0 Six samples were analysed and the following potentials, measured in mV, were obtained; 58.9, 61.0, 60.7, 59.3, 59.6 and 61.7. Use this data to determine the fluoride concentration in the mouthwash in ppm (RAM fluoride 19.0). Comment on whether the levels of fluoride found in the mouthwash conform to the QC specifications.

Appl Of Ms Excel In Analytical Chemistry
2nd Edition
ISBN:9781285686691
Author:Crouch
Publisher:Crouch
Chapter11: Dynamic Electrochemistry
Section: Chapter Questions
Problem 10P
Question
I U
X2 More Formatting
A
A
Q
4. A fluoride ion selective electrode was used to analyse for fluoride as part of the
QC procedures during the production of Listerine Fluoride Defense Anticavity
Mouthwash. The mouthwash contains 0.02% sodium fluoride which is equivalent
to 90 ppm of fluoride. The QC specification for the mouthwash is 90 ± 5 ppm of
fluoride. During production three standards and six periodic samples were
analysed each hour. The mouthwash samples were diluted 100 fold before the
pH and ionic strength were adjusted using an appropriate TISAB before analysis.
The following data was obtained for the standards and the calibration curve is
shown below.
(i)
(ii)
Solution
E (MV)
1.00 x 104 MF
40.5
5.00 x 10-5 MF
1.00 x 10-5 M F
60.2
100.0
Six samples were analysed and the following potentials, measured in mV,
were obtained; 58.9, 61.0, 60.7, 59.3, 59.6 and 61.7. Use this data to
determine the fluoride concentration in the mouthwash in ppm (RAM
fluoride 19.0).
Comment on whether the levels of fluoride found in the mouthwash
conform to the QC specifications.
Transcribed Image Text:I U X2 More Formatting A A Q 4. A fluoride ion selective electrode was used to analyse for fluoride as part of the QC procedures during the production of Listerine Fluoride Defense Anticavity Mouthwash. The mouthwash contains 0.02% sodium fluoride which is equivalent to 90 ppm of fluoride. The QC specification for the mouthwash is 90 ± 5 ppm of fluoride. During production three standards and six periodic samples were analysed each hour. The mouthwash samples were diluted 100 fold before the pH and ionic strength were adjusted using an appropriate TISAB before analysis. The following data was obtained for the standards and the calibration curve is shown below. (i) (ii) Solution E (MV) 1.00 x 104 MF 40.5 5.00 x 10-5 MF 1.00 x 10-5 M F 60.2 100.0 Six samples were analysed and the following potentials, measured in mV, were obtained; 58.9, 61.0, 60.7, 59.3, 59.6 and 61.7. Use this data to determine the fluoride concentration in the mouthwash in ppm (RAM fluoride 19.0). Comment on whether the levels of fluoride found in the mouthwash conform to the QC specifications.
Expert Solution
steps

Step by step

Solved in 2 steps

Blurred answer
Similar questions
  • SEE MORE QUESTIONS
Recommended textbooks for you
Appl Of Ms Excel In Analytical Chemistry
Appl Of Ms Excel In Analytical Chemistry
Chemistry
ISBN:
9781285686691
Author:
Crouch
Publisher:
Cengage
Principles of Instrumental Analysis
Principles of Instrumental Analysis
Chemistry
ISBN:
9781305577213
Author:
Douglas A. Skoog, F. James Holler, Stanley R. Crouch
Publisher:
Cengage Learning
Fundamentals Of Analytical Chemistry
Fundamentals Of Analytical Chemistry
Chemistry
ISBN:
9781285640686
Author:
Skoog
Publisher:
Cengage